Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A, TD-LYTE IMT Standard B + Salt Bridge, Trinidad CH Albumin BCP reagent (Alb_P), Trinidad CH Albumin BCP Calibrator
Electrode, Ion Specific, Sodium. 510(k) K151767, from Siemens Healthcare Diagnostics, Inc. (Newark, DE). Product code JGS, class 2.
As of Nov 17, 2026, K151767 is an FDA 510(k) held by Siemens Healthcare Diagnostics, Inc., substantially equivalent on Dec 11, 2015 after 164 days, with 12 recalls tied to it in FDA's recall database.
- Substantially equivalent: Dec 11, 2015
- Review time: 164 days
- Recalls: 12
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Trinidad CH System is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens. The system’s chemical and immunochemical assay applications utilize photometric and ion selective electrode technology for clinical use.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Bayer Advia 1650 Chemistry SystemK990346 · Bayer Corp. · May 21, 1999
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- A-LYTE® Integrated Multisensor (IMT Na K Cl)K222438 · Siemens Healthcare Diagnostics, Inc. · Jul 25, 2023
- Atellica® CI Analyzer, Atellica® IMThyroid Stimulating Hormone 3-Ultra (TSH3-UL),…K222116 · Siemens Healthcare Diagnostics, Inc. · Jul 13, 2023
- Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica CH Albumin…K161954 · Siemens Healthcare Diagnostics · Mar 9, 2017
Recalls
Recalls FDA ties to K151767, newest first; each opens FDA's recall record| May 22, 2025 | Z-2027-2025Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride. | Open, Classified |
| Dec 15, 2023 | Z-1003-2024Potential for a negative bias with quality control (QC) and patient sample results when using the Atellica CH Immunoglobulin M_2 (IgM_2) reagent. | Open, Classified |
| Nov 18, 2021 | Z-0449-2022(1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay… | Open, Classified |
| May 13, 2021 | Z-2026-2021Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a… | Open, Classified |
| May 4, 2021 | Z-1870-2021Photometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated or depressed patient results | Open, Classified |
| Dec 2, 2020 | Z-0861-2021Multiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient results | Terminated |
| Feb 27, 2020 | Z-1797-2020Editing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the save button on the screen, may corrupt the settings of the assay s test definition, and… | Terminated |
| Oct 11, 2019 | Z-0862-2020On the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and… | Terminated |
| Sep 23, 2019 | Z-0581-2020Potential for calibration error resulting in QC failures after the calibration run with QC and patient results being 5 times higher. | Terminated |
| Jul 30, 2019 | Z-0148-2020Atellica CH 930 Analyzer - Erroneous Indices for Hemolysis and Lipemia in Software versions 1.19.2 and below | Terminated |
| Sep 26, 2018 | Z-0782-2019Multiple Issues Identified in Atellica Solution System Software in V 1.14.2 and lower in: Atellica IM 1300 Analyzer Atellica IM 1600 Analyzer;Atellica CH 930 Analyzer; Atellica Sample Handler Prime | Terminated |
| Jun 29, 2018 | Z-0194-2019Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one… | Terminated |
Other JGS Clearances
The newest 510(k)s for the same product code; their median review was 268 days; see every JGS device- ABL90 FLEX PLUS SystemK252475 · Radiometer Medicals Aps · May 1, 2026
- i-STAT CG8+ cartridge with the i-STAT 1 SystemK230275 · Abbott Point of Care, Inc. · Oct 27, 2023
- A-LYTE® Integrated Multisensor (IMT Na K Cl)K222438 · Siemens Healthcare Diagnostics, Inc. · Jul 25, 2023
- ISE Reagents, Glucose, CRP Latex, DxC 500 AU Clinical Chemistry AnalyzerK220977 · Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd. · Jul 20, 2023
- GLP systems TrackK213486 · Abbott Laboratories · Mar 10, 2022
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 30, 2015
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1665
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files