Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica CH Albumin BCP Reagent (Alb_p), Atellica IM Thyroid Stimulating Hormone TSH, Atellica CH Vancomycin (Vanc)

Electrode, Ion Specific, Sodium. 510(k) K161954, from Siemens Healthcare Diagnostics (Tarrytown, NY). Product code JGS, class 2.

As of Nov 18, 2026, K161954 is an FDA 510(k) held by Siemens Healthcare Diagnostics, substantially equivalent on Mar 9, 2017 after 234 days, with 8 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Mar 9, 2017
  • Review time: 234 days
  • Recalls: 8
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Siemens Atellica™ Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the qualitative and quantitative analysis of various body fluids, using photometric, turbidimetric, chemiluminescent, and integrated ion selective electrode technology for clinical use.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K161954, newest first; each opens FDA's recall record
Oct 21, 2024Z-0496-2025A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or…Open, Classified
Jan 12, 2020Z-1421-2020Some lots ending in "17", "18", "19", or higher may have cuvette defects allowing water from the water bath to contaminate the interior of the cuvette, which may falsely elevate or depress sample results.Open, Classified
Mar 7, 2019Z-1301-2019There were multiple issues identified in the system software which required an update.Terminated
Sep 26, 2018Z-0783-2019Multiple Issues Identified in Atellica Solution System Software in V 1.14.2 and lower in: Atellica IM 1300 Analyzer Atellica IM 1600 Analyzer;Atellica CH 930 Analyzer; Atellica Sample Handler PrimeTerminated
Jun 29, 2018Z-0190-2019Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one…Terminated
Jun 29, 2018Z-0189-2019Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one…Terminated
Jun 29, 2018Z-0191-2019Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one…Terminated
Feb 26, 2018Z-1587-2018When more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzersTerminated

Other JGS Clearances

The newest 510(k)s for the same product code; their median review was 268 days; see every JGS device

Record

Type
Traditional
Received
Jul 18, 2016
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1665

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files