Arca
Skip to content
Tools
Tools

VITROS Chemistry Products Cl- Slides

Electrode, Ion-Specific, Chloride. 510(k) K162020, from Ortho Clinical Diagnostics (Rochester, NY). Product code CGZ, class 2.

As of Nov 18, 2026, K162020 is an FDA 510(k) held by Ortho Clinical Diagnostics, substantially equivalent on Dec 16, 2016 after 148 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 148 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Rx. For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum, plasma, and urine using VITROS Chemistry and Integrated Systems. Chloride measurements are used in diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K162020, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 10, 2019Z-0267-2020Users may experience intermittent slide dispense issues with some cartridges due to the anti-backup platen failing to advance inside the cartridge.Terminated
Jul 11, 2018Z-3184-2018The slides failed to meet current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for urine samples. The measuring range (15-300 mmol/L) for urine samples when tested with the…Terminated

Record

Type
Traditional
Received
Jul 21, 2016
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1170

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files