Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play Module

Pump, Infusion. 510(k) K161469, from Hospira, Inc. (Lake Forest, IL). Product code FRN, class 2.

As of Nov 18, 2026, K161469 is an FDA 510(k) held by Hospira, Inc., substantially equivalent on Mar 28, 2017 after 305 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 305 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Plum 360™ Infusion System with Hospira MedNet™ is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K161469, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 7, 2024Z-2447-2024Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the…Open
Jul 13, 2023Z-2430-2023Due to a manufacturing defect of a supplier provided component, there is a potential that the audible signal for an alarm may not sound under certain conditions.Open
Mar 22, 2023Z-1565-2023Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the…Open
Aug 1, 2018Z-0171-2020A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User…Terminated
Oct 30, 2017Z-0101-2018(1) Under certain conditions, if a malfunction alarm occurs while the pump is in the "Paused" state waiting for the distal pressure to decrease, the pump cannot be turned off and delivery cannot be restarted until the…Terminated
Jun 15, 2017Z-3031-2017Potential for the Connectivity Engine Module of the affected unit to disengage from the main chassis, causing the unit to shut down without an alarm notification. This would delay initiation of therapy or interrupt an…Terminated

Other FRN clearances

The newest 510(k)s for the same product code; their median review was 242 days; see every FRN device

Record

Type
Traditional
Received
May 27, 2016
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files