Plum 360 Infusion System With Hospira Mednet, Smart Card Plug N Play Ce 3.0 Module for Plum A+ Infusion System Upgrade
Pump, Infusion. 510(k) K141789, from Hospira, Inc. (Lake Forest, IL). Product code FRN, class 2.
As of Nov 19, 2026, K141789 is an FDA 510(k) held by Hospira, Inc., substantially equivalent on Jan 16, 2015 after 198 days, with 11 recalls tied to it in FDA's recall database.
- Review time: 198 days
- Recalls: 11
- Cited as a predicate by: 4 devices
Plum 360™ Infusion System with Hospira MedNet™ and Plum A+™ (360 Enabled™) Infusion System are indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.
- Plum A+ Infusion System With Hospira Mednet Software , and Plum A+3 Infusion System With Hospira Mednet SoftwareK042081 · Hospira, Inc. · Aug 24, 2004
- Abbott Plum A+ Infusion PumpK982159 · Abbott Laboratories · Jan 12, 1999
- Lifeshield Latex-Free Microbore Extension Set, Model 14949 and OthersK052722 · Hospira, Inc. · Nov 2, 2005
- Hospira Enteral Feeding Set, Model 20640K061432 · Hospira, Inc. · Sep 11, 2006
- Hospira Infusion Blood SetsK101677 · Hospira, Inc. · Aug 11, 2010
- Sapphire SetsK171346 · Icu Medical · Aug 25, 2017
- Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleK161469 · Hospira, Inc. · Mar 28, 2017
- Hospira Sapphire SetsK160492 · Hospira, Inc. · Dec 6, 2016
- Kangaroo Connect Enteral Feeding Pump with Kangaroo Connect Feeding Sets, Kangaroo Connect PortalK153074 · Covidien · Jul 6, 2016
| May 12, 2026 | Z-3162-2026Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence… | Open |
| May 7, 2024 | Z-2447-2024Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the… | Open |
| Jul 13, 2023 | Z-2430-2023Due to a manufacturing defect of a supplier provided component, there is a potential that the audible signal for an alarm may not sound under certain conditions. | Open |
| Mar 22, 2023 | Z-1565-2023Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the… | Open |
| Jul 29, 2019 | Z-0145-2020, Z-0128-2020, Z-0127-2020Potential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets. | Terminated |
| Jun 15, 2017 | Z-3031-2017Potential for the Connectivity Engine Module of the affected unit to disengage from the main chassis, causing the unit to shut down without an alarm notification. This would delay initiation of therapy or interrupt an… | Terminated |
| Dec 30, 2016 | Z-1682-2017Depleted Battery alarm shows instead of Replace Battery. On battery power, ongoing therapy stops, alarms show and sound, pump shuts down after 3 minutes. On AC power, the pump reboots (takes about 25 seconds). Therapy… | Terminated |
| Mar 25, 2016 | Z-1833-2016, Z-1832-2016Firm's investigation determined that the Proximal Occlusion Alarms observed in the field were attributed to the mechanism calibration process being performed with an out of specification metal cassette, resulting in an… | Terminated |
- RemunityPRO+ Pump for Remodulin® (treprostinil) InjectionK253703 · Deka R&D Corp · Jul 24, 2026
- FreedomEdge Infusion System; High-Flo SubQ Needle Set; Precision Flow Rate TubingK252015 · Koru Medical Systems, Inc. · Jan 28, 2026
- Spectrum IQ Infusion System with Dose IQ Safety Software (3570009)K251636 · Baxter Healthcare Corporation · Jul 28, 2025
- SIGMA Spectrum Infusion Pump with Master Drug LibraryK251640 · Baxter Healthcare Corporation · Jul 28, 2025
- MRidium 3870 MRI Infusion Pump System (3870)K242752 · Iradimed Corporation · May 28, 2025
- BD Alaris Infusion System with Guardrails Suite MXK243855 · Carefusion 303, Inc. · Apr 25, 2025
- Plum Solo™ Precision IV PumpK242114 · Icu Medical, Inc. · Apr 2, 2025
- Plum Duo™ Precision IV PumpK242115 · Icu Medical, Inc. · Apr 2, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 2, 2014
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5725
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files