Arca
Skip to content

Plum Solo™ Precision IV Pump

Pump, Infusion. 510(k) K242114, from Icu Medical, Inc. (Lake Forest, IL). Product code FRN, class 2.

As of Nov 17, 2026, K242114 is an FDA 510(k) held by Icu Medical, Inc., substantially equivalent on Apr 2, 2025 after 257 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Apr 2, 2025
  • Review time: 257 days
  • Recalls: 3

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Plum Solo™ Precision IV Pump is intended for administration of parenteral fluids, medications, and whole blood and blood products through the following routes of administration: intravenous, intra-arterial, and subcutaneous.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K242114, newest first; each opens FDA's recall record
Jun 10, 2026Z-2724-2026In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.Open, Classified
Apr 29, 2026Z-2311-2026Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or…Open, Classified
Dec 15, 2025Z-1120-2026Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at…Open, Classified

Other FRN Clearances

The newest 510(k)s for the same product code; their median review was 215 days; see every FRN device

Record

Type
Traditional
Received
Jul 19, 2024
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.