Pump, Infusion. 510(k) K223607, from Icu Medical, Inc. (Lake Forest, IL). Product code FRN, class 2.
As of Nov 18, 2026, K223607 is an FDA 510(k) held by Icu Medical, Inc., substantially equivalent on Aug 24, 2023 after 265 days, with 6 recalls tied to it in FDA's recall database.
- Review time: 265 days
- Recalls: 6
- Cited as a predicate by: 2 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe Plum Duo Infusion System is intended for parenteral fluids and medications through clinically acceptable routes (limited to intravenous, intra-arterial, and subcutaneous therapies). The Plum Duo Infusion System is intended for use in clinical environments in the hospital environment and other outpatient healthcare facilities by licensed healthcare professionals.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleK161469 · Hospira, Inc. · Mar 28, 2017
Cited as a predicate
2 later 510(k) summaries name this device as a predicate- Plum Duo™ Precision IV PumpK242115 · Icu Medical, Inc. · Apr 2, 2025
- Plum Solo™ Precision IV PumpK242114 · Icu Medical, Inc. · Apr 2, 2025
Recalls
Recalls FDA ties to K223607, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Jun 10, 2026 | Z-2723-2026In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate. | Open |
| Apr 29, 2026 | Z-2312-2026Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or… | Open |
| Dec 15, 2025 | Z-1094-2026ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible… | Open |
| Dec 15, 2025 | Z-1118-2026Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at… | Open |
| Jul 11, 2025 | Z-2129-2025ICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive. | Open |
| Apr 22, 2025 | Z-1990-2025Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager… | Open |
Other FRN clearances
The newest 510(k)s for the same product code; their median review was 242 days; see every FRN device- RemunityPRO+ Pump for Remodulin® (treprostinil) InjectionK253703 · Deka R&D Corp · Jul 24, 2026
- FreedomEdge Infusion System; High-Flo SubQ Needle Set; Precision Flow Rate TubingK252015 · Koru Medical Systems, Inc. · Jan 28, 2026
- Spectrum IQ Infusion System with Dose IQ Safety Software (3570009)K251636 · Baxter Healthcare Corporation · Jul 28, 2025
- SIGMA Spectrum Infusion Pump with Master Drug LibraryK251640 · Baxter Healthcare Corporation · Jul 28, 2025
- MRidium 3870 MRI Infusion Pump System (3870)K242752 · Iradimed Corporation · May 28, 2025
- BD Alaris Infusion System with Guardrails Suite MXK243855 · Carefusion 303, Inc. · Apr 25, 2025
- Plum Solo™ Precision IV PumpK242114 · Icu Medical, Inc. · Apr 2, 2025
- Plum Duo™ Precision IV PumpK242115 · Icu Medical, Inc. · Apr 2, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 2, 2022
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5725
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files