Arca
Skip to content
Tools
Tools

Sapphire Sets

Accessories, Pump, Infusion. 510(k) K171346, from Icu Medical (Lake Forest, IL). Product code MRZ, class 2.

As of Nov 20, 2026, K171346 is an FDA 510(k) held by Icu Medical, substantially equivalent on Aug 25, 2017 after 109 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 109 days
  • Recalls: 9

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Sapphire Primary Sets are indicated for the delivery of fluids from a container to a patient's vascular. Sapphire Blood Sets are indicated for the delivery of fluids including but not limited to blood and blood products from a container to a patient's vascular system. Sapphire Epidural Set are indicated for the delivery of fluids to a patient's epidural space.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K171346, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 29, 2019Z-0135-2020, Z-0129-2020, Z-0132-2020, Z-0130-2020, Z-0133-2020, Z-0131-2020, Z-0147-2020, Z-0134-2020Potential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets.Terminated
Nov 8, 2018Z-0619-2019There have been reports of distal occlusion alarms during the use of SAPPHIRE Epidural Set Yellow-Striped administration sets.Terminated

Other MRZ clearances

The newest 510(k)s for the same product code; their median review was 92 days; see every MRZ device

Record

Type
Traditional
Received
May 8, 2017
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.