Arca
Skip to content
Tools
Tools

ABL800 FLEX with AQURE connectivity

Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K142898, from Radiometer Medical Aps (Dk-Bronshoj, DK). Product code CHL, class 2.

As of Nov 20, 2026, K142898 is an FDA 510(k) held by Radiometer Medical Aps, substantially equivalent on Jun 17, 2015 after 254 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 254 days
  • Recalls: 3
  • Cited as a predicate by: 1 device

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K142898, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 27, 2018Z-1582-2018There is a potential problem relating to the AQURE System, versions 2.3.0 and 2.3.1, that may result in patient mix-up.Terminated
Nov 23, 2016Z-0899-2017Design error when displaying additional information in the patient view window; error may result in misreading a parameter and its value.Terminated
Dec 29, 2015Z-0748-2016The AQURE System has a design error regarding sample type in which sample type may be specified for some results and left blank for other results. The fact that the sample type field is left blank in some cases could…Terminated

Other CHL clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device

Record

Type
Special
Received
Oct 6, 2014
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1120

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.