Arca
Skip to content
Tools
Tools

GEM Premier 7000 with IQM3

Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K223608, from Instrumentation Laboratory Company (Bedford, MA). Product code CHL, class 2.

As of Nov 18, 2026, K223608 is an FDA 510(k) held by Instrumentation Laboratory Company, substantially equivalent on Aug 10, 2023 after 251 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 251 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory.The instrument provides quantitative measurements of pH, pCO2, pO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K223608, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 21, 2026Z-2899-2026The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition…Open
May 8, 2026Z-2822-2026Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.Open

Other CHL clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device

Record

Type
Traditional
Received
Dec 2, 2022
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1120

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files