ABL8X7 Flex

Electrode, Ion Based, Enzymatic, Creatinine. 510(k) K051968, from Radiometer Medical Aps (Bronshoj, DK). Product code CGL, class 2.

As of Nov 19, 2026, K051968 is an FDA 510(k) held by Radiometer Medical Aps, substantially equivalent on Oct 16, 2006 after 453 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 453 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The ABL837 FLEX analyzer is intended for in vitro testing of samples of heparinized whole blood, plasma and serum for the parameter Creatinine. Creatinine measurements measure the concentration of creatinine in blood. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K051968, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 2, 2018Z-0801-2019From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on analyzers could be underestimated with…Terminated
Oct 29, 2014Z-1293-2015Negative drift on QC during in-use lifetime of CREA membranes can be observed.Terminated
Dec 16, 2011Z-1008-2012When the user activates the dosicap (containing the crea additive) and mounts it onto the bottle, the threads will not seal correctly causing approximately 10% of the solution to leak out when the bottle is placed…Terminated
Mar 26, 2009Z-1015-2010Analysis of customer data, (for ABL 8X7 customers) has shown that the reference membranes are out of analytical specifications when replaced as specified in the device labeling. These errors in analytical specifications…Terminated

Other CGL clearances

The newest 510(k)s for the same product code; see every CGL device

Record

Type
Traditional
Received
Jul 20, 2005
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1225

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files