ABL-700 Series, Model Abl 7XX (Various Models)

Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K002290, from Radiometer Medical A/S (Bronshoj, DK). Product code CHL, class 2.

As of Nov 19, 2026, K002290 is an FDA 510(k) held by Radiometer Medical A/S, substantially equivalent on Sep 13, 2000 after 48 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 48 days
  • Recalls: 9
  • Cited as a predicate by: 3 devices

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K002290, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 8, 2012Z-2303-2013RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.Terminated
Jul 5, 2012Z-2163-2012, Z-2162-2012A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is…Terminated
Apr 16, 2009Z-1554-2009When removing the waste bottle from the analyzer waste material may leak both from the analyzer and from the waste bottle.Terminated
Sep 5, 2008Z-0457-2009The firm found that for this particular lot that the barcode, used for entering the characteristics of CAL1 Solution into the analyzer, does not reflect the actual values of the solution. As a consequence, the…Terminated
Jan 18, 2008Z-1653-2008Reported calibration errors are not displayed on the analyzers parameter bar screen or on the patient results when the calibration number is between 32768 - 65535 and 98303 - 131070. This may cause the operator to…Terminated
Nov 13, 2006Z-1094-2007Software anomaly. Protein removing program incorporated in the operating software for Versions: V3.833 and V5.24, of the ABL700 and ABL800 Blood Gas Analyzers, may allow for protein build up on the electrode membranes…Terminated
Sep 29, 2005Z-1386-06Under certain circumstances, the ABL 700 & 800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values…Terminated
Jun 17, 2005Z-1317-06Software defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase.Terminated

Other CHL clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device

Record

Type
Traditional
Received
Jul 27, 2000
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1120

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files