Electrode, Ion Specific, Potassium. 510(k) K092686, from Radiometer Medical Aps (Bronshoj, DK). Product code CEM, class 2.

As of Nov 17, 2026, K092686 is an FDA 510(k) held by Radiometer Medical Aps, substantially equivalent on Aug 6, 2010 after 338 days, with 10 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Aug 6, 2010
  • Review time: 338 days
  • Recalls: 10
  • Cited as a predicate by: 6 devices

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K092686, newest first; each opens FDA's recall record
Mar 21, 2025Z-1659-2025ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software…Open, Classified
Apr 14, 2021Z-1858-2021Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed.Open, Classified
Apr 14, 2021Z-1856-2021Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed.Open, Classified
Feb 3, 2020Z-1760-2020The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a…Terminated
Nov 21, 2019Z-2139-2020The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has…Terminated
Jul 10, 2019Z-2320-2019Software Security; The action is being initiated because of software security vulnerabilities with the firm's analyzer operating system, which may cause the device to shutdown or reboot resulting in delayed medical…Terminated
Jan 26, 2017Z-1462-2017Reports that the sample type "cord blood" has been changed into the two sample types "cord blood arterial" and "cord blood venous.Terminated
Oct 15, 2013Z-0816-2014RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.Terminated
Jan 12, 2011Z-2112-2011Following the receipt of two (2) reported customer complaints, the firm determined that the current operating software for their ABL 90 Analyzer does not prevent the further use of the sensor cassette when there is the…Terminated
Nov 1, 2010Z-1011-2011Due to a software error, a patient mix up can occur on the ABL90. If a sample is pre-registered and the sample "fails" the next result will inherit the patient data from the previous result.Terminated

Other CEM Clearances

The newest 510(k)s for the same product code; their median review was 273 days; see every CEM device

Record

Type
Traditional
Received
Sep 2, 2009
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1600

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.