Arca
Skip to content
Tools
Tools

ABL90 Flex

Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K120197, from Radiometer Medical Aps (Broenshoej, DK). Product code CHL, class 2.

As of Nov 18, 2026, K120197 is an FDA 510(k) held by Radiometer Medical Aps, substantially equivalent on Mar 19, 2012 after 56 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 56 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The ABL90 FLEX is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate and oximetry in heparinized whole blood. The ABL90 FLEX is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate
  • ABL90 FlexK130144 · Radiometer Medical Aps · Jul 31, 2013
  • ABL90 FlexK122729 · Radiometer Medical Aps · Nov 16, 2012
  • ABL90 FlexK122642 · Radiometer Medical Aps · Oct 26, 2012

Recalls

Recalls FDA ties to K120197, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 25, 2018Z-0010-2019The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: The analyzer is in ready mode Operator lifts the inlet and then closes it again, thereby initiating a rinse…Terminated
Dec 8, 2014Z-1046-2015The ABL90 analyzer does not always use the most recent calibration data to calculate patient results. This can in some cases lead to a biased patient result.Terminated
Jul 5, 2012Z-2164-2012A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is…Terminated

Other CHL clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device

Record

Type
Special
Received
Jan 23, 2012
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1120

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files