Duradiagnost
System, X-Ray, Stationary. 510(k) K141381, from Philips Healthcare (Suzhou) Co., Ltd. (Suzhou Jiangsu, CN). Product code KPR, class 2.
As of Nov 18, 2026, K141381 is an FDA 510(k) held by Philips Healthcare (Suzhou) Co., Ltd., substantially equivalent on Jun 12, 2014 after 16 days, with 7 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jun 12, 2014
- Review time: 16 days
- Recalls: 7
- Cited as a predicate by: 2 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe DuraDiagnost is intended for use in generating radiographic images of human anatomy by qualified/trained doctor or technician. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Brivo XR385K103448 · Ge Medical Systems, LLC · Aug 12, 2011
- Philips Bucky VisionK982795 · Philips Medical Systems, Inc. · Nov 24, 1998
- Ddrversa MotionK123005 · Swissray Medical AG · Dec 7, 2012
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- DuraDiagnostK201640 · Philips Healthcare (Suzhou) Co., Ltd. · Jul 9, 2020
- DigitalDiagnost C50K163410 · Philips Healthcare (Suzhou) Co., Ltd. · Jan 4, 2017
Recalls
Recalls FDA ties to K141381, newest first; each opens FDA's recall record| Mar 10, 2025 | Z-1554-2025Potential for collimator to fall as a result of incorrect installation. | Open, Classified |
| Mar 10, 2025 | Z-1553-2025Potential for collimator to fall as a result of incorrect installation. | Open, Classified |
| May 31, 2017 | Z-2322-2017Tube arm assembly could fall down due to cracked welding joints | Terminated |
| Nov 10, 2015 | Z-1696-2016The detector may signal that it is ready for acquisition when it actually is not, resulting in failure to properly acquire the X-ray image. | Terminated |
| Sep 17, 2015 | Z-2682-2016Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable… | Terminated |
| Feb 9, 2015 | Z-1555-2015The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR… | Terminated |
| Jan 23, 2015 | Z-1251-2015Improper installation of Tube arm, and Wall stand may cause it to fall down, Improper installation of table may cause it to float. | Terminated |
Other KPR Clearances
The newest 510(k)s for the same product code; their median review was 195 days; see every KPR device- EXSYS DEXi (EXSYS DEXi-S40S); EXSYS DEXi (EXSYS DEXi-S40P); EXSYS DEXi (EXSYS DEXi-S40T); EXSYS DEXi (EXSYS DEXi-S50HK260227 · DRTECH Corporation · Sep 8, 2026
- MULTIX Impact; MULTIX Impact CK260488 · Siemens Medical Solutions USA, Inc. · Aug 31, 2026
- GC85AK260481 · Samsung Electronics Co., Ltd. · Aug 27, 2026
- DRE DuoK260482 · Jpi Healthcare Co, Ltd. · Jun 15, 2026
- AeroDR TX c02K253446 · Dk Medical Systems Co., Ltd. · Apr 10, 2026
- DRX-Evolution Plus X-ray System, DRX-Compass Mobile X-ray System, Lux HD 35 detector, Lux HD 43 detectorK250954 · Carestream Health · Jan 20, 2026
- uDR Arria & uDR ArisK252000 · Shanghai United Imaging Healthcare Co., Ltd. · Nov 26, 2025
- uDR Aurora CXK251167 · Shanghai United Imaging Healthcare Co., Ltd. · Sep 19, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- May 27, 2014
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1680
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files