Duradiagnost

System, X-Ray, Stationary. 510(k) K141381, from Philips Healthcare (Suzhou) Co., Ltd. (Suzhou Jiangsu, CN). Product code KPR, class 2.

As of Nov 18, 2026, K141381 is an FDA 510(k) held by Philips Healthcare (Suzhou) Co., Ltd., substantially equivalent on Jun 12, 2014 after 16 days, with 7 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jun 12, 2014
  • Review time: 16 days
  • Recalls: 7
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The DuraDiagnost is intended for use in generating radiographic images of human anatomy by qualified/trained doctor or technician. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate
  • DuraDiagnostK201640 · Philips Healthcare (Suzhou) Co., Ltd. · Jul 9, 2020
  • DigitalDiagnost C50K163410 · Philips Healthcare (Suzhou) Co., Ltd. · Jan 4, 2017

Recalls

Recalls FDA ties to K141381, newest first; each opens FDA's recall record
Mar 10, 2025Z-1554-2025Potential for collimator to fall as a result of incorrect installation.Open, Classified
Mar 10, 2025Z-1553-2025Potential for collimator to fall as a result of incorrect installation.Open, Classified
May 31, 2017Z-2322-2017Tube arm assembly could fall down due to cracked welding jointsTerminated
Nov 10, 2015Z-1696-2016The detector may signal that it is ready for acquisition when it actually is not, resulting in failure to properly acquire the X-ray image.Terminated
Sep 17, 2015Z-2682-2016Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable…Terminated
Feb 9, 2015Z-1555-2015The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR…Terminated
Jan 23, 2015Z-1251-2015Improper installation of Tube arm, and Wall stand may cause it to fall down, Improper installation of table may cause it to float.Terminated

Other KPR Clearances

The newest 510(k)s for the same product code; their median review was 195 days; see every KPR device

Record

Type
Traditional, third-party review
Received
May 27, 2014
Review panel
Radiology
Regulation
21 CFR 892.1680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files