Brivo XR385
System, X-Ray, Stationary. 510(k) K103448, from Ge Medical Systems, LLC (Waukesha, WI). Product code KPR, class 2.
As of Nov 18, 2026, K103448 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Aug 12, 2011 after 261 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Aug 12, 2011
- Review time: 261 days
- Recalls: 1
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Brivo XR385 Digital X-Ray System is intended for use on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Silhouette VrK982955 · GE Medical Systems · Nov 4, 1998
- Ge Tethered Portable Digital Radiographic DetectorK041922 · General Electric Co. · Jul 28, 2004
- Canon, Digital Radiography, Model CXDI-50GK031447 · Canon, Inc. · May 21, 2003
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- DuradiagnostK141381 · Philips Healthcare (Suzhou) Co., Ltd. · Jun 12, 2014
Recalls
Recalls FDA ties to K103448, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Sep 16, 2016 | Z-2523-2018The placing of weight on the VRAD detector, of the Brivo XR385 digital radiographic system, can introduce image artifacts, which can result in patient exam retakes. | Terminated |
Other KPR Clearances
The newest 510(k)s for the same product code; their median review was 195 days; see every KPR device- EXSYS DEXi (EXSYS DEXi-S40S); EXSYS DEXi (EXSYS DEXi-S40P); EXSYS DEXi (EXSYS DEXi-S40T); EXSYS DEXi (EXSYS DEXi-S50HK260227 · DRTECH Corporation · Sep 8, 2026
- MULTIX Impact; MULTIX Impact CK260488 · Siemens Medical Solutions USA, Inc. · Aug 31, 2026
- GC85AK260481 · Samsung Electronics Co., Ltd. · Aug 27, 2026
- DRE DuoK260482 · Jpi Healthcare Co, Ltd. · Jun 15, 2026
- AeroDR TX c02K253446 · Dk Medical Systems Co., Ltd. · Apr 10, 2026
- DRX-Evolution Plus X-ray System, DRX-Compass Mobile X-ray System, Lux HD 35 detector, Lux HD 43 detectorK250954 · Carestream Health · Jan 20, 2026
- uDR Arria & uDR ArisK252000 · Shanghai United Imaging Healthcare Co., Ltd. · Nov 26, 2025
- uDR Aurora CXK251167 · Shanghai United Imaging Healthcare Co., Ltd. · Sep 19, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 24, 2010
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1680
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files