Philips Bucky Vision
Solid State X-Ray Imager (Flat Panel/Digital Imager). 510(k) K982795, from Philips Medical Systems, Inc. (Shelton, CT). Product code MQB, class 2.
As of Nov 17, 2026, K982795 is an FDA 510(k) held by Philips Medical Systems, Inc., substantially equivalent on Nov 24, 1998 after 106 days, with 17 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 24, 1998
- Review time: 106 days
- Recalls: 17
- Cited as a predicate by: 11 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Philips BuckyVision is intended for use in general radiographic examinations and applications wherever conventional screen-film systems may be used (excluding fluoroscopy angiography, and mammography).
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Inhealth Blom-Singer Indwelling Valved Insert Accessory DeviceK945287 · Helix Medical, Inc. · Dec 9, 1994
- Sterling Diagnostic Imaging Direct RadiographyK973206 · Sterling Diagnostic Imaging, Inc. · Dec 4, 1997
- Philips ThoravisionK931071 · Philips Medical Systems, Inc. · Sep 15, 1994
Cited as a Predicate
11 later 510(k) summaries name this device as a predicate- DigitalDiagnost C90K182973 · Philips Medical Systems Dmc GmbH · Jan 11, 2019
- DuradiagnostK141381 · Philips Healthcare (Suzhou) Co., Ltd. · Jun 12, 2014
- Philips Pixium 4343RCK131483 · Philips Medical Systems Dmc GmbH · Oct 7, 2013
- SanteviewK111312 · Sehyun Co., Ltd. · Jul 13, 2011
- Wireless Portable Detector FD-W17K090625 · Philips Medical Systems North America Co. · Mar 24, 2009
- DRAD30003/RADREX-IK083503 · Toshiba America Medical Systems, In.C · Feb 13, 2009
- DRAD3000E/RADREX-IK082494 · Toshiba America Medical Systems, In.C · Sep 12, 2008
- DRAD-3000E, Radrex-IK073165 · Toshiba America Medical Systems, In.C · Nov 16, 2007
- Philips Essenta DrK070528 · Philips Medical Systems GmbH, Dmc · Mar 9, 2007
- Philips Xd-S Direct Radiography Workstation/PackageK063781 · Philips Medical Systems North America Co. · Jan 5, 2007
- Paxscan 4030 Medical Digital Imaging SystemK024147 · Varian Medical Systems, Inc. · Feb 12, 2003
Recalls
Recalls FDA ties to K982795, newest first; each opens FDA's recall record| Mar 17, 2014 | Z-1503-2014The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held. | Terminated |
| Aug 8, 2013 | Z-2059-2013The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system. | Terminated |
| Jan 16, 2013 | Z-1448-2013Installation Acceptance Test Document (IATD) non compliant with regulations. Records shows table height measurements out of specification with the consequence of potential system mis-alignment. | Terminated |
| Jan 16, 2013 | Z-1447-2013Installation Acceptance Test Document (IATD) non compliant with regulations IATDs (Installation Acceptance Test could not be located or different templates have been used | Terminated |
| Oct 30, 2012 | Z-0475-2013With Digital Diagnost R2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a… | Terminated |
| Oct 30, 2012 | Z-0900-2013When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction.… | Terminated |
| Oct 25, 2012 | Z-0884-2013When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction.… | Terminated |
| Sep 12, 2012 | Z-2494-2012Hook does not securely hold the footplate in vertical position, causing the footplate to disengage and may fall on patient or operator | Terminated |
| Sep 6, 2012 | Z-2465-2012The Wireless Portable Detector (WPD) is not fixed securely inside the Mobile Detector Holder and may fall down | Terminated |
| Aug 14, 2012 | Z-2384-2012Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm. | Terminated |
| May 22, 2012 | Z-1818-2012With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R… | Terminated |
| Apr 3, 2012 | Z-1464-2012Potential misdiagnosis due to improperly automatically stitched images, improper image ruler handling. | Terminated |
| Nov 1, 2011 | Z-0313-2012Missing a small, metal protective cover on the bottom side of the control handle for the ceiling suspension. exposes printed circuit | Terminated |
| Sep 9, 2010 | Z-0022-2011Potential of broken strands in the main steel cable supporting the accessory holder used with BuckyDiagnost and DigitalDiagnost X-ray systems. | Terminated |
| Jan 21, 2010 | Z-0847-2010Device failed to display appropriate labels indicating Radiation Symbol along with the required wording. | Terminated |
| Dec 13, 2007 | Z-2422-2010When using DICOM print functionality, the printed images may contain data for the wrong patient. | Terminated |
| Jul 23, 2007 | Z-0908-2007Units do not comply with the requirement to provide automatic beam limitation within +/- 3 degrees of horizontal, vertical, and perpendicular to the plane of the image receptor during radiographic procedures. | Terminated |
Other MQB Clearances
The newest 510(k)s for the same product code; their median review was 179 days; see every MQB device- 0909FCE, 0909FCE-HSK253800 · Rayence Co., Ltd. · Jul 31, 2026
- Vivix-S 4386W (FXRD-4386WA)K260820 · Vieworks Co., Ltd. · Jun 18, 2026
- Universal 1417PIK252996 · Konica Minolta Healthcare Americas, Inc. · Jun 12, 2026
- Lux HD 2530 detector (Lux HD 2530)K252911 · Iray Imaging Technology (Haining) Limited · Oct 9, 2025
- Yushan X-Ray Flat Panel DetectorK250211 · Innocare Optoelectronics Corp. · Jul 22, 2025
- Skr 3000K250665 · Konica Minolta, Inc. · Jun 17, 2025
- Wireless/ Wired X-Ray Flat Panel DetectorsK243734 · Allengers Medical Systems Limited · Apr 18, 2025
- Expd 114; Expd 114G; Expd 114P; Expd 114PGK242770 · DRTECH Corporation · Mar 20, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 10, 1998
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1680
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files