Philips Bucky Vision

Solid State X-Ray Imager (Flat Panel/Digital Imager). 510(k) K982795, from Philips Medical Systems, Inc. (Shelton, CT). Product code MQB, class 2.

As of Nov 17, 2026, K982795 is an FDA 510(k) held by Philips Medical Systems, Inc., substantially equivalent on Nov 24, 1998 after 106 days, with 17 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Nov 24, 1998
  • Review time: 106 days
  • Recalls: 17
  • Cited as a predicate by: 11 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Philips BuckyVision is intended for use in general radiographic examinations and applications wherever conventional screen-film systems may be used (excluding fluoroscopy angiography, and mammography).

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

11 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K982795, newest first; each opens FDA's recall record
Mar 17, 2014Z-1503-2014The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held.Terminated
Aug 8, 2013Z-2059-2013The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.Terminated
Jan 16, 2013Z-1448-2013Installation Acceptance Test Document (IATD) non compliant with regulations. Records shows table height measurements out of specification with the consequence of potential system mis-alignment.Terminated
Jan 16, 2013Z-1447-2013Installation Acceptance Test Document (IATD) non compliant with regulations IATDs (Installation Acceptance Test could not be located or different templates have been usedTerminated
Oct 30, 2012Z-0475-2013With Digital Diagnost R2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a…Terminated
Oct 30, 2012Z-0900-2013When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction.…Terminated
Oct 25, 2012Z-0884-2013When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction.…Terminated
Sep 12, 2012Z-2494-2012Hook does not securely hold the footplate in vertical position, causing the footplate to disengage and may fall on patient or operatorTerminated
Sep 6, 2012Z-2465-2012The Wireless Portable Detector (WPD) is not fixed securely inside the Mobile Detector Holder and may fall downTerminated
Aug 14, 2012Z-2384-2012Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.Terminated
May 22, 2012Z-1818-2012With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R…Terminated
Apr 3, 2012Z-1464-2012Potential misdiagnosis due to improperly automatically stitched images, improper image ruler handling.Terminated
Nov 1, 2011Z-0313-2012Missing a small, metal protective cover on the bottom side of the control handle for the ceiling suspension. exposes printed circuitTerminated
Sep 9, 2010Z-0022-2011Potential of broken strands in the main steel cable supporting the accessory holder used with BuckyDiagnost and DigitalDiagnost X-ray systems.Terminated
Jan 21, 2010Z-0847-2010Device failed to display appropriate labels indicating Radiation Symbol along with the required wording.Terminated
Dec 13, 2007Z-2422-2010When using DICOM print functionality, the printed images may contain data for the wrong patient.Terminated
Jul 23, 2007Z-0908-2007Units do not comply with the requirement to provide automatic beam limitation within +/- 3 degrees of horizontal, vertical, and perpendicular to the plane of the image receptor during radiographic procedures.Terminated

Other MQB Clearances

The newest 510(k)s for the same product code; their median review was 179 days; see every MQB device

Record

Type
Traditional
Received
Aug 10, 1998
Review panel
Radiology
Regulation
21 CFR 892.1680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files