Baxter Sigma Spectrum Infusion Pump With Master Drug Library

Pump, Infusion. 510(k) K133801, from Baxter Healthcare Corporation (Round Lake, IL). Product code FRN, class 2.

As of Nov 19, 2026, K133801 is an FDA 510(k) held by Baxter Healthcare Corporation, substantially equivalent on May 7, 2014 after 145 days, with 15 recalls tied to it in FDA's recall database.

  • Review time: 145 days
  • Recalls: 15
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K133801, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 20, 2025Z-2095-2025Certain Spectrum infusion pumps may have an incorrect version of software.Open
Feb 5, 2025Z-1253-2025There is the potential for missing motor mounting screws, which may have occurred during the servicing process.Open
Nov 19, 2024Z-0775-2025Devices were identified as released after repair without full testing being performed, which includes flow testing.Open
Dec 29, 2021Z-0530-2022There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of…Open
Jan 6, 2021Z-1150-2021Battery performance issues related to prolonged storage.Terminated
Nov 23, 2020Z-0714-2021There is a potential software error during programming.Open
Jun 30, 2020Z-0122-2021, Z-0121-2021Baxter is updating the device IFU for safety communication regarding cleaning practices.Terminated
Jun 21, 2018Z-1248-2019The flow rate test verification was not properly executed for certain Sigma Spectrum infusion pumps prior to release. The testing was performed for less than the required minimum run time. Improperly performed flow rate…Terminated
May 8, 2018Z-2113-2018The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However,…Terminated
Oct 6, 2017Z-0971-2018Certain infusion pumps may exhibit excessive wear of the pumping mechanism caused by a lack of lubrication, this could affect the occlusion of the tubing and result in over- or under-infusion.Terminated
Jan 26, 2017Z-1693-2017SIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a Sharp Watchdog Timeout error message. Specifically, the software anomaly causes the error message only under…Terminated
Sep 28, 2016Z-0672-2017A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library and Rear Case Assembly in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an…Terminated
Sep 28, 2016Z-0671-2017A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to…Terminated
Mar 2, 2015Z-1451-2015Loading/Bolus default dose settings in the Master Drug Library and the values shown on the pump during programming may differ. MDL drug dose time in seconds will round to the nearest integer in minutes on the pump dose…Terminated

Other FRN clearances

The newest 510(k)s for the same product code; their median review was 242 days; see every FRN device

Record

Type
Traditional
Received
Dec 13, 2013
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files