Baxter SIGMA Spectrum Infusion Pump with Master Drug Library

Pump, Infusion. 510(k) K230022, from Baxter Healthcare Corporation (Round Lake, IL). Product code FRN, class 2.

As of Nov 20, 2026, K230022 is an FDA 510(k) held by Baxter Healthcare Corporation, substantially equivalent on Mar 31, 2023 after 86 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 86 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K230022, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 20, 2025Z-2096-2025Certain Spectrum infusion pumps may have an incorrect version of software.Open
Nov 19, 2024Z-0776-2025Devices were identified as released after repair without full testing being performed, which includes flow testing.Open
Jul 17, 2024Z-2581-2024The Spectrum WBM may fail to auto-document infusion status information back to the hospital's Electronic Medical Record (EMR) system. This software issue only impacts customers where the Spectrum V8 or Spectrum IQ…Open
Jun 14, 2024Z-2530-2024Improperly performed testing prior to releaseOpen
Jun 15, 2023Z-2103-2023A medical device correction is being issued for SIGMA Spectrum Infusion System (V8 Platform) and Spectrum IQ Infusion System with Dose IQ Safety Software infusion pumps due to an increase in reported false upstream…Open

Other FRN clearances

The newest 510(k)s for the same product code; their median review was 242 days; see every FRN device

Record

Type
Traditional
Received
Jan 4, 2023
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files