Pump, Infusion. 510(k) K042121, from Sigma Intl. (Medina, NY). Product code FRN, class 2.

As of Nov 19, 2026, K042121 is an FDA 510(k) held by Sigma Intl., substantially equivalent on Aug 26, 2004 after 20 days, with 18 recalls tied to it in FDA's recall database.

  • Review time: 20 days
  • Recalls: 18
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Spectrum and Spectrum with Master Drug Library is intended to be used for the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K042121, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 6, 2021Z-1150-2021Battery performance issues related to prolonged storage.Terminated
Nov 23, 2020Z-0712-2021, Z-0713-2021There is a potential software error during programming.Open
Jun 30, 2020Z-0120-2021Baxter is updating the device IFU for safety communication regarding cleaning practices.Terminated
May 8, 2018Z-2112-2018The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However,…Terminated
Apr 3, 2018Z-1665-2018, Z-1667-2018, Z-1666-2018, Z-1664-2018, Z-1663-2018Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line…Terminated
Sep 28, 2016Z-0670-2017A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to…Terminated
Sep 3, 2014Z-2738-2014One Service Technician may not have correctly serviced specific Sigma Spectrum Infusion Pumps according to established procedures during the time period of 5/5/2014 through 6/3/2014.Terminated
Feb 7, 2014Z-1484-2014Baxter Healthcare Corporation has issued an Urgent Device Correction for the SIGMA SPECTRUM Volumetric Infusion Pump with Master Drug Library due to repeated System Error 322 occurrences.Terminated
Sep 24, 2013Z-0189-2014Baxter Healthcare Corporation is recalling the LCD display screen on the SIGMA SPECTRUM Infusion Pump with Master Drug Library for not meeting standards for withstanding an electrostatic discharge.Terminated
Jun 10, 2013Z-0197-2014Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Drug Library due to reports of upstream occlusion alarms and air-in-line alarms occurring during…Terminated
Apr 23, 2012Z-2030-2012Sigma is a safety alert for the Master Drug Library (MDL) Editor software associated with the SIGMA Spectrum Infusion Pump System because it may result in a facility drug library that does not represent common clinical…Terminated
Jul 6, 2011Z-1490-2012Recall expansion; the infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion. Sigma's evaluation…Terminated
Sep 15, 2010Z-0146-2011The infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion.Terminated

Other FRN clearances

The newest 510(k)s for the same product code; their median review was 242 days; see every FRN device

Record

Type
Traditional, third-party review
Received
Aug 6, 2004
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files