Arca
Skip to content
Tools
Tools

V60 Ventilator, Model V8000

Continuous, Minimal Ventilatory Support, Facility Use Ventilator. 510(k) K082660, from Respironics California, Inc. (Carlsbad, CA). Product code MNT, class 2.

As of Nov 20, 2026, K082660 is an FDA 510(k) held by Respironics California, Inc., substantially equivalent on Jan 9, 2009 after 119 days, with 25 recalls tied to it in FDA's recall database.

  • Review time: 119 days
  • Recalls: 25+
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The V60 ventilator is an assist ventilator and is intended to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation: patients with spontaneous respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional settings under the direction of a physician.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K082660, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 6, 2021Z-1137-2021Due to unqualified "design verification" batteries being shipped to customers.Open
Nov 2, 2020Z-0916-2021, Z-0917-2021, Z-0909-2021, Z-0902-2021, Z-0912-2021, Z-0903-2021, Z-0914-2021, Z-0913-2021, Z-0910-2021, Z-0907-2021, Z-0915-2021, Z-0906-2021, Z-0908-2021, Z-0904-2021, Z-0905-2021, Z-0911-2021Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to…Open
Oct 30, 2020Z-0895-2021, Z-0898-2021, Z-0900-2021, Z-0891-2021, Z-0889-2021, Z-0885-2021, Z-0901-2021, Z-0893-2021Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC…Terminated

Other MNT clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every MNT device

Record

Type
Traditional
Received
Sep 12, 2008
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files