Arca
Skip to content
Tools
Tools

Bipap A 30 Ventilatory Support System

Ventilator, Continuous, Non-Life-Supporting. 510(k) K113053, from Respironics, Inc. (Murrysville, PA). Product code MNS, class 2.

As of Nov 19, 2026, K113053 is an FDA 510(k) held by Respironics, Inc., substantially equivalent on Feb 1, 2012 after 111 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 111 days
  • Recalls: 5
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Indications for Use 510(k) Number (if known): _____ Device Name: BiPAP A30 Ventilatory Support System The BiPAP A30 ventilator is intended to provide non-invasive ventilatory support to treat adult and pediatric patients weighing over 10 kg (22lbs) with Obstructive Sleep Apnea (OSA and Respiratory Insufficiency.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K113053, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 9, 2025Z-2036-2025This device does not indicate for use in patients with respiratory failure.Open
Aug 26, 2022Z-1654-2022Motor assemblies manufactured with nonconforming plastic that contributes to off-gassing of VOCs.Open
Jun 14, 2021Z-1957-2021, Z-1958-2021The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.Open
Jun 14, 2021Z-1973-2021The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the device s the air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.Open

Other MNS clearances

The newest 510(k)s for the same product code; their median review was 269 days; see every MNS device

Record

Type
Abbreviated
Received
Oct 13, 2011
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files