Ventilator, Continuous, Facility Use. 510(k) K981072, from Respironics, Inc. (Pittsburgh, PA). Product code CBK, class 2.
As of Nov 19, 2026, K981072 is an FDA 510(k) held by Respironics, Inc., substantially equivalent - subject to tracking reg. on Nov 12, 1998 after 233 days, with 11 recalls tied to it in FDA's recall database.
- Review time: 233 days
- Recalls: 11
- Cited as a predicate by: 10 devices
The ESPRIT ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult and pediatric patients as prescribed by a physician. The ESPRIT Ventilator is intended for use in either invasive or non-invasive applications.
- V200 Ventilator With Aprv Mode, Esprit Ventilator With Aprv ModeK110083 · Respironics California, Inc. · Jan 13, 2012
- V200 VentilatorK102054 · Respironics California, Inc. · Dec 22, 2010
- V60 Ventilator, Model V8000K082660 · Respironics California, Inc. · Jan 9, 2009
- Esprit Ventilator Auto-Trak Sensitivity Option, Model V1000K072450 · Respironics California, Inc. · Dec 12, 2007
- Esprit Ventilator Speaking Mode Option, Model V1000K071212 · Respironics California, Inc. · Jul 30, 2007
- Esprit Ventilator With Nico-Esprit Interface Option, Model V1000K041412 · Respironics California, Inc. · Jun 18, 2004
- Esprit Ventilator, Model V1000K034040 · Respironics, Inc. · Jan 21, 2004
- Plv Continuum Ventilator, Model P2000K034032 · Respironics California, Inc. · Mar 16, 2004
- Bird Tbird Vela VentilatorK032451 · Bird Products Corp. · Mar 17, 2004
- Plv Continuum Ventilator, Model P1000K022750 · Respironics California, Inc. · Nov 20, 2003
| Oct 29, 2014 | Z-0485-2015If the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the ventilator to switch… | Terminated |
| Sep 17, 2014 | Z-0008-2015A failure of a specific component may prevent the ventilators from operating on AC power or transitioning back to AC power when operating on battery. Additionally, if a battery is not present or is depleted the… | Terminated |
| Apr 22, 2009 | Z-1813-2009The recall was initiated because through routine product monitoring Respironics discovered that under certain conditions, when operating on AC line power, a portion of the current from the External Battery charger is… | Terminated |
| Jan 25, 2008 | Z-2002-2008, Z-2003-2008, Z-2004-2008This action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in… | Terminated |
| Oct 29, 2007 | Z-2084-2013The power to the Esprit Display's Backlight is interrupted, causing the GUI to flicker or go dark and making the display unreadable. | Terminated |
| Aug 8, 2006 | Z-0047-2007This action is being initiated due to power supply snubber board failures on certain Esprit Ventilators in countries that utilize operating voltages of 200-240 Volts AC and will involve the replacement of the Esprit… | Terminated |
| Feb 20, 2002 | Z-1230-04Flow sensors would exhibit drift. | Terminated |
| Jun 8, 2001 | Z-0951-03Material hardness on original design check valve may cause premature failure. | Terminated |
| Jun 1, 2001 | Z-0903-04Ventilator would spontaneously suspend ventilatory support to patients due to a hardware/software compatibility issue. | Terminated |
- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
- Type
- Traditional
- Received
- Mar 24, 1998
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files