Plv Continuum Ventilator, Model P1000
Continuous, Ventilator, Home Use. 510(k) K022750, from Respironics California, Inc. (Carlsbad, CA). Product code NOU, class 2.
As of Nov 19, 2026, K022750 is an FDA 510(k) held by Respironics California, Inc., substantially equivalent on Nov 20, 2003 after 458 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 458 days
- Recalls: 2
- Cited as a predicate by: 1 device
The intended use of the PLV Continuum ventilator is to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The intended patient population includes pediatric and adult patients who weigh at least 5 kg (11 lbs).
- Portable Lifecare Ventilator PLV-102K842876 · Lifecare Services, Inc. · Aug 29, 1984
- Esprit VentilatorK981072 · Respironics, Inc. · Nov 12, 1998
- Ltv 1000K984056 · Pulmonetic Systems, Inc. · Apr 12, 1999
- Plv Continuum Ventilator, Model P2000K034032 · Respironics California, Inc. · Mar 16, 2004
| Mar 20, 2006 | Z-0808-06Flow valve failures- When failure occurs, air flow from the ventilator ceases resulting in the PLVC I transitioning into a 'Vent Inop' mode whereby the safety valve is activated, opening the circuit to ambient air.… | Terminated |
| Nov 18, 2005 | Z-0354-06During product testing of the PLV Continuum Ventilator's Safety Valve/Pressure Relief Valve (SV/PRV) at our manufacturing facility, Respironics found that when the PLY Continuum Ventilator is operated in the 'Face Up'… | Terminated |
- Vivo 45 LSK233452 · Breas Medical AB · Jan 13, 2025
- Vivo 45 LSK193586 · Breas Medical AB · Jan 27, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 19, 2002
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files