Ltv 1000
Ventilator, Continuous, Facility Use. 510(k) K984056, from Pulmonetic Systems, Inc. (Colton, CA). Product code CBK, class 2.
As of Nov 18, 2026, K984056 is an FDA 510(k) held by Pulmonetic Systems, Inc., substantially equivalent on Apr 12, 1999 after 150 days, with 12 recalls tied to it in FDA's recall database.
- Review time: 150 days
- Recalls: 12
- Cited as a predicate by: 9 devices
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Tbird Vs/Avs Volume Ventilator, Models 15568, 15570 , LP10 Volume Ventilator With Pressure Limit Model 00-004130, Bear 3K981971 · Bird Products Corp. · Aug 24, 1998
- Model 7700 Quantum Pressure Support Ventilator PsvK962517 · Healthdyne Technologies, Inc. · Feb 20, 1997
Cited as a predicate
9 later 510(k) summaries name this device as a predicate- Newport HT70 Family of VentilatorsK090888 · Newport Medical Instruments, Inc. · Nov 18, 2009
- Newport HTS50 Ventilator With Dual Pac Battery SystemK082724 · Newport Medical Instruments, Inc. · Dec 16, 2008
- Carina System, Model 8419350; Carina Basic Device, Model 5704110K072885 · Draeger Medical AG & Co. KG · Apr 7, 2008
- Palmtop Ventilator, Models PTV-8, PTV-10K070594 · Pulmonetic Systems, Inc. · Aug 1, 2007
- Ltv 1000 VentilatorK051767 · Pulmonetic Systems, Inc. · Jul 29, 2005
- Modification To Ltv 1000 Ventilator/Breathing CircuitsK040790 · Pulmonetic Systems, Inc. · Jun 3, 2004
- Modification To Ltv 1000 VentilatorK040540 · Pulmonetic Systems, Inc. · Apr 29, 2004
- Plv Continuum Ventilator, Model P2000K034032 · Respironics California, Inc. · Mar 16, 2004
- Plv Continuum Ventilator, Model P1000K022750 · Respironics California, Inc. · Nov 20, 2003
Recalls
Recalls FDA ties to K984056, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Feb 22, 2008 | Z-1463-2008, Z-1469-2008, Z-1468-2008, Z-1467-2008, Z-1464-2008, Z-1466-2008, Z-1465-2008, Z-1470-2008The failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver. | Terminated |
| Nov 9, 2005 | Z-0186-06, Z-0184-06, Z-0185-06Pulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the… | Terminated |
| Sep 8, 2004 | Z-1485-04Ventilators have not consistently switched to internal battery operation when the external DC power source has become inadequate to supply stable power. This situation results in loss of ventilation to the patient. | Terminated |
Other CBK clearances
The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 13, 1998
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files