Heartstart Mrx, Models M3535A Or M3536A
Automated External Defibrillators (Non-Wearable). 510(k) K063375, from Philips Medical Systems (Andover, MA). Product code MKJ, class 3.
As of Nov 17, 2026, K063375 is an FDA 510(k) held by Philips Medical Systems, substantially equivalent on Jan 11, 2007 after 64 days, with 11 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jan 11, 2007
- Review time: 64 days
- Recalls: 11
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Heartstart Mrx, Model M3535AK031187 · Philips Medical Systems · Aug 27, 2003
- Heartstart Mrx With Q-Cpr Option, Models M3535A Or M3536AK051134 · Philips Medical Systems · Sep 8, 2005
- Heartstart Mrx, Invasive Pressure & Temperature OptionsK061707 · Philips Medical Systems · Aug 11, 2006
Recalls
Recalls FDA ties to K063375, newest first; each opens FDA's recall record| Apr 10, 2020 | Z-2064-2020Damaged ESD bags used for storing components in Processor PCA Replacement Kits | Terminated |
| Feb 26, 2020 | Z-1499-2020Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical… | Terminated |
| Jan 14, 2020 | Z-1342-2020M4758A Ambulance Parameter Extension (APE) kit may cause ECG "Leads Off" condition in hot, humid environments | Terminated |
| Jul 16, 2019 | Z-2488-2019Non-conforming devices are identified, which may not have been included in prior field actions and may not have had a number of corrections applicable to these devices. | Terminated |
| Jun 7, 2019 | Z-1923-2019The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring and failure or… | Terminated |
| May 3, 2018 | Z-2035-2018The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. Failure to charge could potentially cause therapy to be… | Terminated |
| Sep 28, 2017 | Z-0278-2018Philips MRx devices have an EtCO2 module that may contain contamination; this contamination (will/may) cause the device to be unable to display the EtCO2 or CO2 waveform (capnogram). | Terminated |
| Apr 14, 2016 | Z-1612-2016MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. | Terminated |
| Dec 23, 2014 | Z-0219-2016MRx monitor/defibrillator could reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed | Terminated |
| Nov 19, 2014 | Z-0204-20161. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery… | Terminated |
| Nov 19, 2014 | Z-0963-2015Device may malfunction, which could cause therapy to be delayed, disabled, or delivered inadvertently. | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Nov 8, 2006
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.5310
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files