Heartstart Mrx, Model M3535A
Automated External Defibrillators (Non-Wearable). 510(k) K031187, from Philips Medical Systems (Andover, MA). Product code MKJ, class 3.
As of Nov 17, 2026, K031187 is an FDA 510(k) held by Philips Medical Systems, substantially equivalent on Aug 27, 2003 after 134 days, with 21 recalls tied to it in FDA's recall database.
- Substantially equivalent: Aug 27, 2003
- Review time: 134 days
- Recalls: 21
- Cited as a predicate by: 5 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.
Cited as a Predicate
5 later 510(k) summaries name this device as a predicate- Philips Heartstart Xl+, Philips Heartstart External Paddles, Philips Multifunction…K110825 · Philips Medical Systems · Feb 16, 2012
- Intellivue Cl SP02 Pod Model 865215, Intellivue Cl Nbp Pod Model 865216, Intellivue…K101600 · Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · Aug 31, 2010
- Heartstart Mrx, Models M3535A Or M3536AK063375 · Philips Medical Systems · Jan 11, 2007
- Heartstart Mrx Monitor/DefibrillatorK062233 · Philips Medical Systems North America, Inc. · Nov 22, 2006
- Heartstart Mrx, Invasive Pressure & Temperature OptionsK061707 · Philips Medical Systems · Aug 11, 2006
Recalls
Recalls FDA ties to K031187, newest first; each opens FDA's recall record| Sep 14, 2020 | Z-0285-2021The periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has created an addendum… | Terminated |
| Sep 14, 2020 | Z-0284-2021The periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has created an addendum… | Terminated |
| Jul 16, 2020 | Z-2794-2020The therapy switch may fail, resulting in the device exhibiting the following behaviors: the device may not perform the selected function, the therapy knob may not change to the energy setting selected, or the device… | Terminated |
| Sep 28, 2017 | Z-0278-2018Philips MRx devices have an EtCO2 module that may contain contamination; this contamination (will/may) cause the device to be unable to display the EtCO2 or CO2 waveform (capnogram). | Terminated |
| Jul 17, 2017 | Z-0348-2018Issue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator… | Terminated |
| Jul 12, 2017 | Z-0489-2018Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode. | Terminated |
| Feb 24, 2017 | Z-1417-2017HeartStart MRx Monitor/Defibrillator may may exhibit abnormal behavior when operating using a single battery as the sole source of power may fail to deliver therapy or monitoring | Terminated |
| Aug 26, 2016 | Z-2842-2016The handle can separate from the MRx housing due to breakage of mounts on the rear case. | Terminated |
| Apr 14, 2016 | Z-1612-2016MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. | Terminated |
| Oct 14, 2015 | Z-0320-2016The following MRx software issue has been identified: MRx model M3535A with software version F.03.06 and earlier, and model M3536A with version T.00.05 and earlier may stop the automated Ready-For-Use (RFU) test in an… | Terminated |
| Dec 23, 2014 | Z-0219-2016MRx monitor/defibrillator could reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed | Terminated |
| Nov 19, 2014 | Z-0262-2016MRx Defib can be susceptible to one or both issues: 1. The C02 Inlet Port associated with end-tidal carbon dioxide (EtCO2) monitoring on MRx Monitor/Defibrillators can be pushed into the MRx housing, making it… | Terminated |
| Nov 19, 2014 | Z-0273-2016When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast… | Terminated |
| Nov 19, 2014 | Z-0204-20161. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery… | Terminated |
| Nov 19, 2014 | Z-0963-2015Device may malfunction, which could cause therapy to be delayed, disabled, or delivered inadvertently. | Terminated |
| Jan 19, 2010 | Z-1135-2010Paddle shock switch may fail to actuate and delay or prevent delivery of defibrillation therapy | Terminated |
| Aug 21, 2009 | Z-0003-2010incorrect default configuration settings, insufficient battery power as the user may not be alerted that no battery is installed or that a low battery condition exists prior to disconnecting from AC or DC power | Terminated |
| Jul 7, 2009 | Z-1795-2009External transport use environments, the mechanical/electrical connection between the pads therapy cable including pads/CPR cable and the MRx/MRxE may wear at an increased rate ultimately prevent the device from sensing… | Terminated |
| Jul 14, 2008 | Z-2337-2008Defective internal memory cards may exhibit one or more of the following behaviors: "Device repeatedly restarts approximately every 20 seconds "Slow device start-up of approximately 15 seconds or more. During this time,… | Terminated |
| Apr 29, 2005 | Z-0848-05The device may disarm and not deliver shock when performing a Synchronized Cardioversion | Terminated |
| Dec 15, 2004 | Z-0341-05In Software versions below A.02.00, with the sync function enabled and the device switched from manual to AED, there may be an unexpected delay in delivery of energy | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Abbreviated
- Received
- Apr 15, 2003
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.5310
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files