Arca
Skip to content
Tools
Tools

Heartstart Mrx With Q-Cpr Option, Models M3535A Or M3536A

Automated External Defibrillators (Non-Wearable). 510(k) K051134, from Philips Medical Systems (North Attleboro, MA). Product code MKJ, class 3.

As of Nov 19, 2026, K051134 is an FDA 510(k) held by Philips Medical Systems, substantially equivalent on Sep 8, 2005 after 128 days, with 25 recalls tied to it in FDA's recall database.

  • Review time: 128 days
  • Recalls: 25+
  • Cited as a predicate by: 5 devices

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K051134, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 27, 2020Z-2157-2020HeartStart XL MonitorlDefibrillators may not have been included in prior field actions and received required correctionsTerminated
Nov 23, 2018Z-0603-2019Affected lithium-ion batteries may contain a defective component (Thermal Cut-off or TCO). Should the component fail, the battery will no longer charge or deliver power to the Monitor/Defibrillator, which may prevent…Terminated
Jan 29, 2018Z-0684-2018Inform users of correct placement of the disposable adhesive pad on the QCPR meter and inform users of injuries associated with proper performance of CPR. As required by Consent Decree of Permanent Injunction No.…Terminated
Sep 28, 2017Z-0278-2018Philips MRx devices have an EtCO2 module that may contain contamination; this contamination (will/may) cause the device to be unable to display the EtCO2 or CO2 waveform (capnogram).Terminated
Jul 17, 2017Z-0348-2018Issue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator…Terminated
Jul 12, 2017Z-0489-2018Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.Terminated
Feb 24, 2017Z-1417-2017HeartStart MRx Monitor/Defibrillator may may exhibit abnormal behavior when operating using a single battery as the sole source of power may fail to deliver therapy or monitoringTerminated
Aug 26, 2016Z-2842-2016The handle can separate from the MRx housing due to breakage of mounts on the rear case.Terminated
Apr 14, 2016Z-1612-2016MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component.Terminated
Oct 14, 2015Z-0320-2016The following MRx software issue has been identified: MRx model M3535A with software version F.03.06 and earlier, and model M3536A with version T.00.05 and earlier may stop the automated Ready-For-Use (RFU) test in an…Terminated
Dec 23, 2014Z-0219-2016MRx monitor/defibrillator could reboot at an indeterminate time, potentially causing therapy to be interrupted or delayedTerminated
Nov 19, 2014Z-0262-2016MRx Defib can be susceptible to one or both issues: 1. The C02 Inlet Port associated with end-tidal carbon dioxide (EtCO2) monitoring on MRx Monitor/Defibrillators can be pushed into the MRx housing, making it…Terminated
Nov 19, 2014Z-0273-2016When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast…Terminated
Nov 19, 2014Z-0204-20161. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery…Terminated
Nov 19, 2014Z-0963-2015Device may malfunction, which could cause therapy to be delayed, disabled, or delivered inadvertently.Terminated
Feb 18, 2014Z-1090-2014A component of the MRx Processor Board may be susceptible to damage from electrostatic discharge (ESD), which can disrupt ECG and SpO2 functionality leading to an inability to perform.Terminated
Feb 18, 2014Z-1120-2014Philips HeartStart MRx Monitor/Defibrillator, when operating on battery power only, may experience an unexpected shutdown if exposed to elevated levels of electromagnetic interference from RF energy sourceTerminated
Feb 18, 2014Z-1194-2014When the HeartStart MRx.is used with the Q-CPR Meter in defibrillation mode, the Q-CPR Meter may incorrectly display the Do Not Touch the Patient icon.Terminated
Feb 18, 2014Z-1108-2014Wireless Link may be unable to transmit data when configured for a specific data flow.Terminated
Feb 18, 2014Z-1118-2014ECG trunk cable and connector block of the MRx could be susceptible to accelerated wear, which could result in an interrupted ECG signalTerminated
Feb 14, 2014Z-1109-2014Philips HeartStart XL may experience a power board failure causing failure to defibrillateTerminated
Feb 4, 2014Z-1110-2014HeartStart MRx Defibrillator Monitor therapy connection could experience accelerated wear causing risk for patients and/or caregiversTerminated
Dec 10, 2013Z-0570-2014MRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHgTerminated
Jun 19, 2013Z-1797-2013The MRx may deliver a non-synchronized cardioversion shock resulting in the delivery of incorrect therapy, which may induce ventricular fibrillation.Terminated
Jun 7, 2013Z-0593-2014Error in the printing of the therapy energy setting labels on Philips HeartStart M3536A MRx Monitor/Defibrillators. The first energy setting is labeled 10 when it should be labeled1-10Terminated

Record

Type
Traditional
Received
May 3, 2005
Review panel
Cardiovascular
Regulation
21 CFR 870.5310

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.