Medley Syringe Pump Module System, Model 8110

Pump, Infusion. 510(k) K023264, from Alaris Medical Systems, Inc. (San Diego, CA). Product code FRN, class 2.

As of Nov 20, 2026, K023264 is an FDA 510(k) held by Alaris Medical Systems, Inc., substantially equivalent on Dec 19, 2002 after 80 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 80 days
  • Recalls: 7
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The MEDLEY™ Syringe Pump is intended for use in today’s growing professional healthcare environment for facilities that utilize syringe pumps for the delivery of fluids, medications, blood and blood products.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K023264, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 11, 2026Z-1645-2026Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care UnitsOpen
Nov 17, 2017Z-0322-2018The syringe plunger grippers may fail to automatically close around the syringe plunger press when the gripper control knob has been closed.Terminated
Aug 8, 2016Z-2879-2016A software anomaly with the Alaris Syringe module software version 9.15 may cause an infusion to unexpectedly stop when the Syringe module is transitioning from one rate to another.Terminated
Jul 21, 2015Z-2362-2015Channel Error code is displayed on the PC unit with an audio and visual alarm, and on the syringe module. After the error is cleared on the PCU, the syringe pump is unresponsive to key presses until the next power…Terminated
Jun 12, 2009Z-1794-2009There is a potential risk of IUI connector failure in the Alaris System when these connectors become corroded or contaminated, which may result in an interruption of infusion with an audible and visual alarm. IUI…Terminated
Mar 5, 2008Z-1712-2008A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors due to failure of the U9 Socketed Integrated Circuits (U9 IC) on the…Terminated
Mar 5, 2008Z-1711-2008A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors with the Alaris System due to the Inter-Unit Interface (IUI) connectors…Terminated

Other FRN clearances

The newest 510(k)s for the same product code; their median review was 242 days; see every FRN device

Record

Type
Traditional
Received
Sep 30, 2002
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files