Accutni and Accutni Calibrators On The Access Immunoassay Systems, Model #33340, 33345

Immunoassay Method, Troponin Subunit. 510(k) K021814, from Beckman Coulter, Inc. (Chaska, MN). Product code MMI, class 2.

As of Nov 18, 2026, K021814 is an FDA 510(k) held by Beckman Coulter, Inc., substantially equivalent on Sep 27, 2002 after 116 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 116 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Access AccuTnI assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma using the Access Immunoassay Systems to aid in the diagnosis and treatment of myocardial infarction and cardiac muscle damage.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K021814, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 11, 2010Z-1995-2010The recall was initiated because as a follow up to PCA # 12605 dated December 16, 2009, Beckman Coulter has confirmed additional reports that each of the following assays may produce lower than expected results when run…Terminated
Feb 6, 2010Z-1318-2010The recall was initiated because Beckman Coulter has confirmed customer reports that different results have been obtained using the same patient samples on Access/Access 2 and UniCel Dxl platforms. Values obtained with…Terminated
Dec 10, 2009Z-1883-2011The recall was initiated after Beckman Coulter confirmed that under certain circumstances, the following assays may produce elevated values at the low end when run in conjunction with the enhanced Access Accu Tnl assay…Terminated

Other MMI clearances

The newest 510(k)s for the same product code; their median review was 266 days; see every MMI device

Record

Type
Traditional
Received
Jun 3, 2002
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1215

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files