Arca
Skip to content
Tools
Tools

Architect Stat Troponin-I Immunoassay

Immunoassay Method, Troponin Subunit. 510(k) K041192, from Fisher Diagnostics (Middletown, VA). Product code MMI, class 2.

As of Nov 20, 2026, K041192 is an FDA 510(k) held by Fisher Diagnostics, substantially equivalent on Aug 12, 2004 after 98 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 98 days
  • Recalls: 4

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

ARCHITECT STAT Troponin-I is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of cardiac troponin-I (cTnI) in human serum and plasma on the ARCHITECT i System with STAT protocol capability. Troponin-I values are used to assist in the diagnosis of myocardial infarction (MI).

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K041192, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 26, 2022Z-1660-2022One lot of reagent was manufactured with insufficient quantities of the heterophilic blocking agents in the conjugate component.Open
Apr 11, 2012Z-1509-2012The lot of Troponin-I is demonstrating a shift in expected results in some cases, and varies from kit to kit. A concentration change of up to +/-55% could occur when switching kits without recalibrating.Terminated
Mar 8, 2007Z-0855-2007False Negative Results : The analytical sensitivity claim of less than or equal to 0.01 ng/mL (ug/L) might not be met for all lots of Architect STAT Troponin-I reagents, thus false negative results may occur with…Terminated
Dec 21, 2005Z-0513-06There has been an increase in customer complaints observed for failure to calibrate or inability to generate patient result errors.Terminated

Other MMI clearances

The newest 510(k)s for the same product code; their median review was 266 days; see every MMI device

Record

Type
Traditional
Received
May 6, 2004
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1215

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files