Accutni and Accutni Calibrators On The Access Immunoassay System # 33340, 33345
Immunoassay Method, Troponin Subunit. 510(k) K010429, from Beckman Coulter, Inc. (Chaska, MN). Product code MMI, class 2.
As of Nov 20, 2026, K010429 is an FDA 510(k) held by Beckman Coulter, Inc., substantially equivalent on Jun 5, 2001 after 112 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 112 days
- Recalls: 3
- Cited as a predicate by: 1 device
- Accutni and Accutni Calibrators On The Access Immunoassay Systems, Model #33340, 33345K021814 · Beckman Coulter, Inc. · Sep 27, 2002
| Mar 11, 2010 | Z-1995-2010The recall was initiated because as a follow up to PCA # 12605 dated December 16, 2009, Beckman Coulter has confirmed additional reports that each of the following assays may produce lower than expected results when run… | Terminated |
| Feb 6, 2010 | Z-1318-2010The recall was initiated because Beckman Coulter has confirmed customer reports that different results have been obtained using the same patient samples on Access/Access 2 and UniCel Dxl platforms. Values obtained with… | Terminated |
| Dec 10, 2009 | Z-1883-2011The recall was initiated after Beckman Coulter confirmed that under certain circumstances, the following assays may produce elevated values at the low end when run in conjunction with the enhanced Access Accu Tnl assay… | Terminated |
- i-STAT hs-TnI cartridge with the i-STAT Alinity SystemK253051 · Abbott Point of Care, Inc. · Jun 11, 2026
- VITROS Immunodiagnostic Products hs Troponin I Reagent PackK252393 · Ortho-Clinical Diagnostics, Inc. · Oct 29, 2025
- Access hsTnIK243483 · Beckman Coulter, Inc. · Aug 1, 2025
- Access hsTnIK242870 · Beckman Coulter, Inc. · Jun 16, 2025
- i-STAT hs-TnI cartridge with the i-STAT 1 SystemK240984 · Abbott Point of Care, Inc. · Jan 3, 2025
- Atellica® IM High-Sensitivity Troponin I (TnIH)K241165 · Siemens Healthcare Diagnostics, Inc. · Jul 25, 2024
- PATHFAST®hs-cTnI-IIK231974 · Phc Corporation · Mar 20, 2024
- Access hsTnIK222881 · Beckman Coulter, Inc. · Dec 18, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 13, 2001
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1215
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files