Arca
Skip to content

Elecsys Troponin I And Troponin I Stat Test Systems

Immunoassay Method, Troponin Subunit. 510(k) K082699, from Roche Diagnostics (Indianapolis, IN). Product code MMI, class 2.

As of Nov 17, 2026, K082699 is an FDA 510(k) held by Roche Diagnostics, substantially equivalent on Aug 19, 2009 after 338 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Aug 19, 2009
  • Review time: 338 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Elecsys Troponin I Immunoassay: Immunoassay for the in vitro quantitative determination of cardiac troponin I in human serum and plasma. The Elecsys Troponin I assay is intended to aid in the diagnosis of myocardial infarction. The electrochemiluminescence immunoassay “ECLIA” is intended for use on the Elecsys and MODULAR Analytics E170 immunoassay analyzers.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K082699, newest first; each opens FDA's recall record
Apr 7, 2021Z-1595-2021Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and…Open, Classified
Apr 7, 2021Z-1596-2021Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and…Open, Classified
Mar 12, 2012Z-1388-2012An investigation has confirmed a considerably reduced recovery of Li-heparin plasma samples for Troponin I lot 163176 and Troponin I STAT lot 163177. In the worst case, Li-Heparin sample recovery may be as low as 50% of…Terminated
Mar 12, 2012Z-1387-2012An investigation has confirmed a considerably reduced recovery of Li-heparin plasma samples for Troponin I lot 163176 and Troponin I STAT lot 163177. In the worst case, Li-Heparin sample recovery may be as low as 50% of…Terminated

Other MMI Clearances

The newest 510(k)s for the same product code; their median review was 266 days; see every MMI device

Record

Type
Traditional
Received
Sep 15, 2008
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1215

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files