Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K013562, from Baxter Healthcare Corp (Mcgaw Park, IL). Product code KDI, class 2.

As of Nov 27, 2026, K013562 is an FDA 510(k) held by Baxter Healthcare Corp, substantially equivalent on Dec 12, 2002 after 412 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 412 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The AUORA TM System 1000 ® Series Single Patient Dialysis Delivery System is part of a high permeability hemodialysis system which consists of a controlled dialysate delivery system that incorporates an ultrafiltration controller to prevent excessive loss of water from the patient's blood, an extracorporeal blood set, and a high permeability dialyzer.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K013562, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 31, 2009Z-0179-2010When the main power is lost temporarily, a software anomaly prevents the instrument from automatically changing the ultrafiltration rate (UFR) to the minimum value as described in the Operator's manual.Terminated
Nov 18, 2005Z-0347-06Potential for fluid or air to be passed through the venous line clamp if the tubing is not centered in the clamp and extends beyond the edge of the clamp's pinch zone. This could result in an air emboli condition.Terminated

Other KDI clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every KDI device

Record

Type
Traditional
Received
Oct 26, 2001
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.