Drake Willock System 1000 Dialysate Delivery Syst
System, Dialysate Delivery, Single Patient. 510(k) K910215, from Cd Medical, Inc. (Miami Lakes, FL). Product code FKP, class 2.
As of Nov 27, 2026, K910215 is an FDA 510(k) held by Cd Medical, Inc., substantially equivalent on Apr 9, 1991 after 82 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 82 days
- Recalls: 5
- Cited as a predicate by: 4 devices
- Drake Willock System & Altra Touch 1000 Dialysis Delivery SystemK970446 · Althin Medical AB an Affiliate of Baxter Intl · Jun 2, 1997
- Drake Willock System 1000 Dialysate Delivery System Altratouch 1000 Dialysate Delivery SystemK964922 · Althin Medical AB an Affiliate of Baxter Intl · May 21, 1997
- Altratouch 1000 Hemodialysis Delivery SystemK954987 · Althin Medical AB an Affiliate of Baxter Intl · Aug 8, 1996
- Aurora System 1000 Series Home Self-Care Single Patient Delivery System; Models- Sys 1000, L3 With N100 Arm, 4B SoftwareK013562 · Baxter Healthcare Corp · Dec 12, 2002
| Jun 21, 2006 | Z-1253-06The UF Removal Regulators were assembled incorrectly. The mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy. | Terminated |
| Feb 24, 2006 | Z-0766-06The older style 'Through Hardware' Dialysate Monitoring Manifolds and 'Noryl' Ultra Filtration Removal Regulators may become cracked due to mechanical and thermal stress. If either of these parts become cracked during… | Terminated |
| Nov 18, 2005 | Z-0347-06Potential for fluid or air to be passed through the venous line clamp if the tubing is not centered in the clamp and extends beyond the edge of the clamp's pinch zone. This could result in an air emboli condition. | Terminated |
| May 28, 2003 | Z-0395-05The air detector may not detect air bubbles consistently at the selected limit. | Terminated |
| Mar 20, 2003 | Z-0833-03The dialysis treatment may be run without the bicarbonate due to an undetected bicarbonate pump stall following a sodium change. | Terminated |
- Type
- Traditional
- Received
- Jan 17, 1991
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.5820
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files