Drake Willock System & Altra Touch 1000 Dialysis Delivery System

Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K970446, from Althin Medical AB an Affiliate of Baxter Intl (Portland, OR). Product code KDI, class 2.

As of Nov 27, 2026, K970446 is an FDA 510(k) held by Althin Medical AB an Affiliate of Baxter Intl, substantially equivalent on Jun 2, 1997 after 117 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 117 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Indications for Use of the System 1000® - The intended use of the device is to provide hemodialysis treatments in the acute and chronic setting, including high flux hemodialysis. - The device is intended to be used by trained operators when prescribed by a physician. - The device is intended to be used in conjunction with a hollow fiber or parallel plate dialyzer.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K970446, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 19, 2011Z-0173-2012The Tina System was released to the customer with OOS values for the Pressure Displays Accuracy and Leakage Rate Test.Terminated
Jun 21, 2006Z-1253-06The UF Removal Regulators were assembled incorrectly. The mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy.Terminated
Jun 14, 2005Z-1394-05Operator's Manuals addenda are being issued to provide correct information concerning the correct syringe size to use for heparin infusion, the use of chemical disinfectants in the System 1000, TMP Advise, and…Terminated
Dec 14, 2004Z-0367-05Air bubbles have been observed in the tubing sets past the air detector of the instrument with no alarms. Air bubbles in the circuit can cause an air embolism.Terminated
Feb 20, 2004Z-0583-04These devices have the potential to overheat in the absence of a Ground Fault Circuit Interrupter (GFCI). This heater system failure could result in a fire.Terminated

Other KDI clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every KDI device

Record

Type
Traditional
Received
Feb 5, 1997
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files