Alaris Auto ID Module Model 8600

Recall Z-2723-2020, part of FDA recall event 85691. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class I device recall on Jun 30, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 22, 2024.

  • Classification: Class I
  • Started: Jun 30, 2020
  • Status: Terminated
  • Products in this event: 8

Why It Was Recalled

Class I

Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.

Record

Product
Alaris Auto ID Module Model 8600, a modular infusion pump and monitoring system
Quantity
17767
Lots and codes
During cleaning, updated cleaning instructions and IUI covers must be used for all serial numbers.
Distribution
Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, and BH
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Aug 6, 2020

Other Products in This Recall

Recall event 85691

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.