Alaris System PC Unit Model 8000
Recall Z-2716-2020, part of FDA recall event 85691. See CareFusion 303's company page.
CareFusion 303, Inc. started this Class I device recall on Jun 30, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 22, 2024.
- Classification: Class I
- Started: Jun 30, 2020
- Status: Terminated
- Products in this event: 8
Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.
- Product
- Alaris System PC Unit Model 8000, a modular infusion pump and monitoring system
- Quantity
- 37141
- Lots and codes
- During cleaning, updated cleaning instructions and IUI covers must be used for all serial numbers. ** US Serial numbers: 2326370, 2326382, 2326385, 2326387, 3668143, 2326400, 2326401, 2326402, 2326416, 2326418, 2326419, 2326423, 2326647, 2326664, 2326819, 2326821, 2326834, 2326837, 2326842, 2326873, 2326882, 2326883, 2326926, 9300064, 9300081, 9300083, 9300117, 9300149, 9300162, 9300163, 9300164, 9300171, 9300174, 9300176, 9300284, 9300302, 9300313, 9300317, 9300333, 9300334, 9300485, 3669844, 9300531, 9300532, 9300533, 9300534, 9300535, 9300538, 3686560, 4011408, 3696229, 3000933, 4122102,…
- Distribution
- Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, and BH
- Recalling firm
- CareFusion 303, Inc., San Diego, CA, United States
- How the firm notified
- Letter
- FDA classified
- Aug 6, 2020
- Alaris System PC Unit Model 8015, a modular infusion pump and monitoring systemZ-2717-2020 · Class I · Jun 30, 2020 · Terminated
- Alaris Pump Module Model 8100, a modular infusion pump and monitoring systemZ-2718-2020 · Class I · Jun 30, 2020 · Terminated
- Alaris EtCO2 Module Model 8300, a modular infusion pump and monitoring systemZ-2721-2020 · Class I · Jun 30, 2020 · Terminated
- Alaris Auto ID Module Model 8600, a modular infusion pump and monitoring systemZ-2723-2020 · Class I · Jun 30, 2020 · Terminated
- Alaris SpO2 Module Model 8210 and Model 8220, a modular infusion pump and monitoring systemZ-2722-2020 · Class I · Jun 30, 2020 · Terminated
- Alaris PCA Module Model 8120, a modular infusion pump and monitoring systemZ-2720-2020 · Class I · Jun 30, 2020 · Terminated
- Alaris Syringe Module Model 8110, a modular infusion pump and monitoring systemZ-2719-2020 · Class I · Jun 30, 2020 · Terminated
Newest first, one product per recall event; search them all
- BD Pyxis MedStation ES Tower REF: 352 Medication cabinetZ-2302-2026 · Class II · Apr 30, 2026 · Ongoing
- BD Alaris" System with Guardrails" Suite MX with Point of Care UnitZ-1645-2026 · Class II · Feb 11, 2026 · Ongoing
- BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03,…Z-1028-2026 · Class II · Dec 3, 2025 · Ongoing
- BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES…Z-0962-2026 · Class II · Nov 20, 2025 · Ongoing
- BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a…Z-1217-2026 · Class II · Nov 6, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports