Arca
Skip to content
Tools
Tools

BD Alaris" System with Guardrails" Suite MX with Point of Care Unit

Recall Z-1645-2026, part of FDA recall event 98250. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class II device recall on Feb 11, 2026 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Feb 11, 2026
  • Status: Ongoing

Why It Was Recalled

Class II

Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units

Record

Product
BD Alaris" System with Guardrails" Suite MX with Point of Care Unit
Quantity
1,025,567
Lots and codes
Catalog Number: 8015 UDI-DI codes: 10885403210747 10885403210754 10885403210891 10885403212048 10885403222108 10885403224614 10885403515231 10885403516023 10885403519215 10885403519284 10885403519291 10885403801518 10885403801532 10885403801549 Serial Numbers: 12360968 12360983 12360985 12360986 12360987 12360988 12360989 12360990 12360991 12360992 12360993 12360994 12360995 12360996 12360997 12360998 12360999 12361000 12361001 12361002 12361003 12361004 12361005 12361006 12361008 12361010 12361013 12361017 12361018 12361019 12361020 12361021 12361022 12361024 12361026 12361027 12361028…
Distribution
Worldwide distribution: U.S.(Nationwide) to states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP. MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY O.U.S. (foreign) to countries of: Australia, Belgium, Bahrain, Botswana, Canada, China, Cuba, Germany, India, Israel, Italy, Japan, Korea, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Pakistan, Philippines, Qatar, Saudi Arabia, Singapore, South Africa, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Mar 26, 2026

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports