3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack
Recall Z-2640-2023, part of FDA recall event 92915. See 3M Company - Health Care Business's company page.
3M Company - Health Care Business started this Class II device recall on Aug 22, 2023 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Aug 22, 2023
- Status: Ongoing
- Products in this event: 4
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.
- Product
- 3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air removal steam sterilization processes
- Quantity
- 77,280 devices
- Lots and codes
- UDI/DI: 50707387787283( (shipper); Lot Codes: 41482VF, exp. 33P4ND; 33P4YN, exp. 01/02/2025
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.
- Recalling firm
- 3M Company - Health Care Business, Saint Paul, MN, United States
- How the firm notified
- N/A
- FDA classified
- Sep 26, 2023
- 3M Attest Super Rapid Readout Biological Indicator, REF 1492V; used to qualify or monitor dynamic-air-removal steam…Z-2637-2023 · Class II · Aug 22, 2023 · Ongoing
- 3M Attest Super Rapid Readout Steam Challenge Pack, REF 1496V; designed to qualify and monitor dynamic-air removal…Z-2638-2023 · Class II · Aug 22, 2023 · Ongoing
- 3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482V; designed to qualify and monitor dynamic-air…Z-2639-2023 · Class II · Aug 22, 2023 · Ongoing
- Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 …Z-0743-2025 · Class II · Nov 22, 2024 · Ongoing
- 3M Tegaderm CHG, Chlorhexidine Gluconate I.V. Securement Dressing, REF 1658RZ-0038-2025 · Class II · Aug 14, 2024 · Ongoing
- 3M Durapore Surgical Tape, Catalog Number 1538-118Z-1633-2024 · Class II · Mar 20, 2024 · Ongoing
- 3M Tube Securement Device, Small, 1.7 IN/PO x 2.4 IN/PO, REF 1500U, Sterile; and 3M Tube Securement Device, Large, 2.2…Z-1475-2024 · Class II · Feb 20, 2024 · Ongoing
- 3M Specialty Blade Assembly, REF 9660,70-2011-8491-1, 7100213026, used with used with 3M Surgical Clipper Catalog…Z-1170-2024 · Class II · Jan 12, 2024 · Ongoing
- 3M Attest Steam Chemical Integrators, REF 1243AZ-1609-2023 · Class II · Apr 6, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports