3M Specialty Blade Assembly
Recall Z-1170-2024, part of FDA recall event 93825. See 3M Company - Health Care Business's company page.
3M Company - Health Care Business started this Class II device recall on Jan 12, 2024 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Jan 12, 2024
- Status: Ongoing
- Products in this event: 2
3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41¿C (105.8¿F). " Do not leave the clipper blade applied to the same location on the patient s skin for longer than 1 second as potential injury may occur; continuous motion of the clipper during use is required. Note: If clipper is left running unused in air for 10 minutes without applying to skin, the Model 9690 blade maximum temperature may reach 60¿C (140¿F) and the Model 9660 blade maximum temperature may reach 77¿C (170.6¿F)."
- Product
- 3M Specialty Blade Assembly, REF 9660,70-2011-8491-1, 7100213026, used with used with 3M Surgical Clipper Catalog Number 9661L
- Quantity
- 764,580 blades
- Lots and codes
- UDI/DI 50707387789294 (shipper), 30707387789290 (tray), all lot codes
- Distribution
- Worldwide distribution.
- Recalling firm
- 3M Company - Health Care Business, Saint Paul, MN, United States
- How the firm notified
- Letter
- FDA classified
- Feb 21, 2024
- 3M Specialty Blade Assembly, REF 9660,70-2011-8462-2, 7100213114, used with used with 3M Surgical Clipper Catalog…Z-1169-2024 · Class II · Jan 12, 2024 · Ongoing
- Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 …Z-0743-2025 · Class II · Nov 22, 2024 · Ongoing
- 3M Tegaderm CHG, Chlorhexidine Gluconate I.V. Securement Dressing, REF 1658RZ-0038-2025 · Class II · Aug 14, 2024 · Ongoing
- 3M Durapore Surgical Tape, Catalog Number 1538-118Z-1633-2024 · Class II · Mar 20, 2024 · Ongoing
- 3M Tube Securement Device, Small, 1.7 IN/PO x 2.4 IN/PO, REF 1500U, Sterile; and 3M Tube Securement Device, Large, 2.2…Z-1475-2024 · Class II · Feb 20, 2024 · Ongoing
- 3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482V; designed to qualify and monitor dynamic-air…Z-2639-2023 · Class II · Aug 22, 2023 · Ongoing
- 3M Attest Steam Chemical Integrators, REF 1243AZ-1609-2023 · Class II · Apr 6, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports