3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack
Recall Z-2639-2023, part of FDA recall event 92915. See 3M Company - Health Care Business's company page.
3M Company - Health Care Business started this Class II device recall on Aug 22, 2023 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Aug 22, 2023
- Status: Ongoing
- Products in this event: 4
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.
- Product
- 3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482V; designed to qualify and monitor dynamic-air removal steam sterilization processes
- Quantity
- 172,464 devices
- Lots and codes
- UDI/DI: 50707387787276( (shipper); Lot Codes: 33P3EF, exp. 12/27/2024; 33P4ND, exp. 01/01/2025; 33P4NR, exp. 01/01/2025; 33P5CA, exp. 01/03/2025; 33P5XT, exp. 01/06/2025
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.
- Recalling firm
- 3M Company - Health Care Business, Saint Paul, MN, United States
- How the firm notified
- N/A
- FDA classified
- Sep 26, 2023
- 3M Attest Super Rapid Readout Biological Indicator, REF 1492V; used to qualify or monitor dynamic-air-removal steam…Z-2637-2023 · Class II · Aug 22, 2023 · Ongoing
- 3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air…Z-2640-2023 · Class II · Aug 22, 2023 · Ongoing
- 3M Attest Super Rapid Readout Steam Challenge Pack, REF 1496V; designed to qualify and monitor dynamic-air removal…Z-2638-2023 · Class II · Aug 22, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports