iLet Bionic Pancreas
Recall Z-2513-2026, part of FDA recall event 98517. See Beta Bionics's company page.
Beta Bionics, Inc. started this Class II device recall on Sep 23, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Sep 23, 2025
- Status: Ongoing
The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.
- Product
- iLet Bionic Pancreas, REF: BB1001
- Quantity
- 4580
- Lots and codes
- iLet ACE Pump firmware BOM000072 Rev C (version 3.3.6) and BOM000072 Rev B (version 3.3.4)
- Distribution
- US: WI, HI, OH, ME, CO, NJ, NE, KY, SC, CA, TX, NH, NY, WA, PA, MD, FL, IA, GA, MI, VT, TN, IL, IN, KS, MA, LA, MS, NC, CT, MN, VA, OK, AR, AZ, MO, MT, RI, DC, NM, OR, ID, NV, AK, ND, DE, WV, WY, SD, UT, AL
- Recalling firm
- Beta Bionics, Inc., Irvine, CA, United States
- How the firm notified
- Letter
- FDA classified
- Jun 18, 2026
- iLet Bionic Pancreas, REF: BB1001Z-1996-2026 · Class II · Mar 12, 2026 · Ongoing
- iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision SoftwareZ-1809-2026 · Class II · Aug 13, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports