iLet Bionic Pancreas
Recall Z-1996-2026, part of FDA recall event 98475. See Beta Bionics's company page.
Beta Bionics, Inc. started this Class II device recall on Mar 12, 2026 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Mar 12, 2026
- Status: Ongoing
insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.
- Product
- iLet Bionic Pancreas, REF: BB1001
- Quantity
- 1,080
- Lots and codes
- Software versions 1.4.3, and 1.4.2
- Distribution
- US Nationwide distribution in the states of TX, PA, AL, MI, OH, FL, TN, GA.
- Recalling firm
- Beta Bionics, Inc., Irvine, CA, United States
- How the firm notified
- —
- FDA classified
- Apr 29, 2026
- iLet Bionic Pancreas, REF: BB1001Z-2513-2026 · Class II · Sep 23, 2025 · Ongoing
- iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision SoftwareZ-1809-2026 · Class II · Aug 13, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports