iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software
Recall Z-1809-2026, part of FDA recall event 98528. See Beta Bionics's company page.
Beta Bionics, Inc. started this Class II device recall on Aug 13, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Aug 13, 2025
- Status: Ongoing
it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.
- Product
- iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software
- Quantity
- 15 units
- Lots and codes
- ILet Ace Pumps containing software version BOM000036 Rev C (version 1.4.2) or older.
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, MA, MD, ME, MI, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WY AND WASHINGTON DC.
- Recalling firm
- Beta Bionics, Inc., Irvine, CA, United States
- How the firm notified
- Letter
- FDA classified
- Apr 10, 2026
- iLet Bionic Pancreas, REF: BB1001Z-1996-2026 · Class II · Mar 12, 2026 · Ongoing
- iLet Bionic Pancreas, REF: BB1001Z-2513-2026 · Class II · Sep 23, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports