Nitinol Staple, Elastic Staple Large-Asymmetrical
Recall Z-2262-2018, part of FDA recall event 80099. See TriMed's company page.
TriMed Inc. started this Class II device recall on Apr 13, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Jan 22, 2021.
- Classification: Class II
- Started: Apr 13, 2018
- Status: Terminated
- Products in this event: 4
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
- Product
- Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19
- Quantity
- 532
- Lots and codes
- Lot #: a) R901, 03UU, 050522, 10505-22, 0344, 0051, 11805, 00AJ, 10505-23, 050523, 180502, 180901, 180503, 18902; b) 03UV, 050523, 10505-23, 00AJ, 0052, 180901, 11809, 180902, 26408, 264-08, 11805, 180502, 118901
- Distribution
- U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden
- Recalling firm
- TriMed Inc., Santa Clarita, CA, United States
- How the firm notified
- Letter
- FDA classified
- Jun 20, 2018
- Nitinol Staple, Elastic Staple Small-Asymmetrical, Sterile R, Model #: a) ES-10x13x15; b) ES-10x16x19; c) ES-10x17x19;…Z-2264-2018 · Class II · Apr 13, 2018 · Terminated
- Nitinol Staple, Elastic Staple Large-Symmetrical, Sterile R, Model #: a) ES-18x14; b) ES-20x20; c) ES-25x22Z-2263-2018 · Class II · Apr 13, 2018 · Terminated
- Nitinol Staple, Elastic Staple Small-Symmetrical, Sterile R, Model #: a) ES-08x08; b) ES-10x10; c) ES-10x13; d)…Z-2265-2018 · Class II · Apr 13, 2018 · Terminated
- TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm,…Z-1791-2024 · Class II · Mar 19, 2024 · Ongoing
- Volar Bearing Plate. REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S, VBEAL-3-7W, VBEAL-5-7W,…Z-0932-2024 · Class II · Nov 28, 2023 · Ongoing
- RipCord Syndesmosis Button, REF: STA001KZ-1101-2023 · Class II · Dec 22, 2022 · Ongoing
- TriMed Wrist Hook Dorsal Impactor, REF IMPCT-WHD; and TriMed Wrist Hook Volar Impactor, REF IMPCT-WHV, both nonsterile,…Z-0359-2020 · Class II · Sep 28, 2018 · Ongoing
- TriMed Dorsal Wrist Hook Plate, Dorsal, 4-Hole, REF WHD-4, contained in a tray, insert, or poly package containing…Z-0118-2020 · Class II · Aug 24, 2018 · Terminated
- TriMed Hex Cortical Screw 4.0, 40 mm, REF HEX4.0-40, non-Sterile, RX, a component of the TriMed Ankle Fixation System. …Z-1214-2019 · Class II · Aug 24, 2018 · Terminated
- TriMed Olecranon Hook Plate, Lt, 6-Holes, REF: OHOOKL-6, Non-Sterile, RX only, (01) 00842188101281 - Product Usage:…Z-0366-2020 · Class II · Aug 15, 2018 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports