RipCord Syndesmosis Button
Recall Z-1101-2023, part of FDA recall event 91478. See TriMed's company page.
TriMed Inc. started this Class II device recall on Dec 22, 2022 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Dec 22, 2022
- Status: Ongoing
Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia, can exceed its strength causing it to break.
- Product
- RipCord Syndesmosis Button, REF: STA001K
- Quantity
- 76
- Lots and codes
- UDI-DI: 00842188123849, Lot: 85305-01
- Distribution
- US Nationwide distribution in the states of IL, PA, TX, GA, TN, AL, MO, FL, AR, LA, HI, OH, TX, KS
- Recalling firm
- TriMed Inc., Santa Clarita, CA, United States
- How the firm notified
- Letter
- FDA classified
- Feb 7, 2023
Newest first, one product per recall event; search them all
- TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm,…Z-1791-2024 · Class II · Mar 19, 2024 · Ongoing
- Volar Bearing Plate. REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S, VBEAL-3-7W, VBEAL-5-7W,…Z-0932-2024 · Class II · Nov 28, 2023 · Ongoing
- TriMed Wrist Hook Dorsal Impactor, REF IMPCT-WHD; and TriMed Wrist Hook Volar Impactor, REF IMPCT-WHV, both nonsterile,…Z-0359-2020 · Class II · Sep 28, 2018 · Ongoing
- TriMed Dorsal Wrist Hook Plate, Dorsal, 4-Hole, REF WHD-4, contained in a tray, insert, or poly package containing…Z-0118-2020 · Class II · Aug 24, 2018 · Terminated
- TriMed Hex Cortical Screw 4.0, 40 mm, REF HEX4.0-40, non-Sterile, RX, a component of the TriMed Ankle Fixation System. …Z-1214-2019 · Class II · Aug 24, 2018 · Terminated
- TriMed Olecranon Hook Plate, Lt, 6-Holes, REF: OHOOKL-6, Non-Sterile, RX only, (01) 00842188101281 - Product Usage:…Z-0366-2020 · Class II · Aug 15, 2018 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports