TriMed Olecranon Hook Plate
Recall Z-0366-2020, part of FDA recall event 83984. See TriMed's company page.
TriMed Inc. started this Class II device recall on Aug 15, 2018 (voluntary: firm initiated); FDA lists it as terminated.
- Classification: Class II
- Started: Aug 15, 2018
- Status: Terminated
The firm has become aware that the angled-hole of certain semi-tubular design bone plates may have been modified from its specification, and the modification may not have been fully qualified.
- Product
- TriMed Olecranon Hook Plate, Lt, 6-Holes, REF: OHOOKL-6, Non-Sterile, RX only, (01) 00842188101281 - Product Usage: The OHOOKL is a semi-tubular design bone plate with two offset intramedullary tines, which are intended to penetrate the small terminal fragment to provide additional stabilization.
- Quantity
- 583 Bone Plates
- Lots and codes
- Lot # 110631,
- Distribution
- US Nationwide distribution in the states of MD, UT.
- Recalling firm
- TriMed Inc., Santa Clarita, CA, United States
- How the firm notified
- Letter
- FDA classified
- Nov 13, 2019
Newest first, one product per recall event; search them all
- TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm,…Z-1791-2024 · Class II · Mar 19, 2024 · Ongoing
- Volar Bearing Plate. REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S, VBEAL-3-7W, VBEAL-5-7W,…Z-0932-2024 · Class II · Nov 28, 2023 · Ongoing
- RipCord Syndesmosis Button, REF: STA001KZ-1101-2023 · Class II · Dec 22, 2022 · Ongoing
- TriMed Wrist Hook Dorsal Impactor, REF IMPCT-WHD; and TriMed Wrist Hook Volar Impactor, REF IMPCT-WHV, both nonsterile,…Z-0359-2020 · Class II · Sep 28, 2018 · Ongoing
- TriMed Dorsal Wrist Hook Plate, Dorsal, 4-Hole, REF WHD-4, contained in a tray, insert, or poly package containing…Z-0118-2020 · Class II · Aug 24, 2018 · Terminated
- TriMed Hex Cortical Screw 4.0, 40 mm, REF HEX4.0-40, non-Sterile, RX, a component of the TriMed Ankle Fixation System. …Z-1214-2019 · Class II · Aug 24, 2018 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports