Arca
Skip to content
Tools
Tools

SpaceLabs Arkon Anesthesia Delivery System with Version 2.0 Software installed. This…

Recall Z-1442-2014, part of FDA recall event 67781. See Del Mar Reynolds Medical's company page.

Del Mar Reynolds Medical, Ltd. started this Class I device recall on Mar 10, 2014 (voluntary: firm initiated); FDA lists it as terminated, terminated Aug 20, 2014.

  • Classification: Class I
  • Started: Mar 10, 2014
  • Status: Terminated

Why It Was Recalled

Class I

The System may go into a controlled failed state if a series of parameters on the user interface are activated in a specific sequence.

Record

Product
SpaceLabs Arkon Anesthesia Delivery System with Version 2.0 Software installed. This system includes a ventilator.
Quantity
16 units distributed in the US
Lots and codes
SERIAL NUMBERS OF UNITS DISTRIBUTED IN THE US: ARKN-000011, ARKN-000016, ARKN-000017, ARKN-000019, ARKN-000020, ARKN-000021, ARKN-000022, ARKN-000023, ARKN-000024, ARKN-000025, ARKN-000026, ARKN-000027, ARKN-000028, ARKN-000029, ARKN-000030, and ARKN-000031.
Distribution
USA Nationwide Distribution in the states of North Carolina and South Carolina
Recalling firm
Del Mar Reynolds Medical, Ltd., Hertford, N/A, United Kingdom
How the firm notified
Letter
FDA classified
Apr 11, 2014

Del Mar Reynolds Medical, Ltd.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.