Spacelabs BleaseSirius Anesthesia Workstation

Recall Z-0968-2013, part of FDA recall event 64535. See Del Mar Reynolds Medical's company page.

Del Mar Reynolds Medical, Ltd. started this Class I device recall on Jan 15, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated May 14, 2013.

  • Classification: Class I
  • Started: Jan 15, 2013
  • Status: Terminated

Why It Was Recalled

Class I

A defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being delivered to the patient.

Record

Product
Spacelabs BleaseSirius Anesthesia Workstation, BleaseFocus Anesthesia Workstation, and CAS I/II Absorber contained in Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation installed with the CAS I/II Absorber. Anesthesia workstation for use in the hospital environment and locations not requiring portability.
Quantity
108 total (17 units in the US and 91 units outside the US)
Lots and codes
serial numbers: ABSO-004448, ABSO-004458, ABSO-004459, ABSO-004460, ABSO-004461, ABSO-004462, ABSO-004463, ABSO-004465, ABSO-004466, ABSO-004467, ABSO-004468, ABSO-004469, ABSO-004470, ABSO-004471, ABSO-004472, ABSO-004473, ABSO-004474, ABSO-004475, ABSO-004476, ABSO-004477, ABSO-004478, ABSO-004479, ABSO-004480, ABSO-004481, ABSO-004482, ABSO-004483, ABSO-004487, ABSO-004488, ABSO-004489, ABSO-004490, ABSO-004491, ABSO-004492, ABSO-004493, ABSO-004494, ABSO-004495, ABSO-004497, ABSO-004499, ABSO-004500, ABSO-004503, ABSO-004504, ABSO-004505, ABSO-004506, ABSO-004507, ABSO-004508,…
Distribution
Worldwide Distribution - USA Nationwide including the states of New Mexico, North Carolina and the countries of Argentina, China, Mexico, and Taiwan.
Recalling firm
Del Mar Reynolds Medical, Ltd., Hertford, N/A, United Kingdom
How the firm notified
Letter
FDA classified
Mar 20, 2013